Maintenance, cleaning, servicing, disposal
CAUTION
The devices are subject to an annual safety and functional test (maintenance) in
accordance with §7 of the Medical Devices Operator Ordinance "Maintenance of
Medical Devices".
With the optional accessory "smart assist ISO", a metrological test (MTP) is carried out
at intervals of 2 years, in accordance with the manufacturer's specifications.
A selection of suitable products for maintenance can be found at
WARNING
The safety and effectiveness of the equipment can be optimally maintained if the
care and maintenance intervals are observed.
Component
Maintenance
interval
Testing and maintenance
Display
daily
Only use cleaning and disinfection agents suitable for the
display. Do not apply the agent directly, but use a clean
microfiber cloth.
Cushion
after training
Only use the disinfectants approved by the
manufacturer to disinfect the upholstery.
A list of possible disinfectants can be found at
https://www.servicehelper.de/maintenance/.
Cladding
Floormats
Frame parts
regular
Clean with warm water and a microfiber cloth. Do not
use solvents, chlorides, polishes, chemical cleaners or
wax polishes.
Screws
monthly
Check externally accessible screws for tightness.
Guide rods
monthly
Check for sufficient lubrication and lubricate with suitable
maintenance oil.
Castors
monthly
Check for sufficient lubrication.