Page 40 of 68
Proton Products InteliSENS DG1130 Diameter Gauge Instruction Manual (Issue 1a)
Continuous Data Output
The continuous data output mode continuously outputs measurement data approximately every
100ms.
Continuous data output is enabled by sending the following ASCII string to the gauge:
H<CR>
Continuous data output is disabled by sending the following ASCII string to the gauge:
I<CR>
where
<CR>
is the Carriage Return character.
When continuous data output mode is enabled, the gauge will continuously output measurement data
approximately every 100ms as an ASCII string formatted as follows:
$(Gauge type)(Diameter)(Status)(Position in gate)<CR><LF>(Units)(Axis)
RS-422
/
RS-485
C
OMMUNICATIONS
This RS-422 / RS-485 interface is fitted as standard. It operates independently of the other
communications interfaces and may be accessed at the same time as them.
RS-422 / RS-485 interface
The RS-422 / RS-485 interface may be accessed through the “RS232/485” connector.
Connector type: DB9 female (socket)
Pin
RS-422
Designation
Comment
RS-485 two wire operation
Link
Link
Designation
1
RS422_A
A+
4
RS422_B
B-
5
GND_R
GND
6
RS422_Y
9
RS422_Z
S
Shield
Ensure that the cable shield is connected to this
via the plug shield connection.
Shield
The four-wire RS-422 interface may be converted to two-wire RS-485 operation by installation of the 2
shorting links shown above.