4.4.2.1 Check of motion-release function
•
Depress the motion-release button (Fig. 2-2/6) without pressing any other of the motion buttons
(Fig. 2-2/1, 2, 3 or 7, 8).
No motion may start.
•
Depress one of the motion buttons (Fig. 2-2/1, 2 or 7, 8) without pressing the motion-release
button (Fig. 2-2/6).
No motion may start.
If table height motion starts during this check, put the table out of order and call service.
4.4.3 Regular Checks
4.4.3.1 Quality Control
X-ray equipment should be quality-controlled at regular intervals or as required by
applicable regulations to determine that the image quality remains in accordance to national
regulations, e.g. by a monthly consistency testing.
4.4.4 Technical safety checks and maintenance
Required technical safety checks and maintenance (see “Maintenance check list” in the
manual “Technical Description”) must be performed at 24-months intervals by
PROTEC
Service or authorized service providers to ensure the safe and reliable operation of the
equipment.
In the event that scheduled maintenance is not performed, PROTEC will not be responsible
for damages occurred by the user or third parties if such damages are the result of
improper or missing maintenance.
4.4.5 Disposal Remarks
The accumulator-boxes of the PROGNOST XPE-Akku, all mechanical, electrical and plastic
components have to be disposed according to local or national regulations. In case of doubt contact
PROTEC.