PROGNOST XPE-Akku, Instruction for Use
5033-0-0002
30/10/2019
PROTEC GmbH & Co. KG
,
In den Dorfwiesen 14,
71720 Oberstenfeld, Germany
7 of 33
(Fig. 2-2/6).
Status 2, 4 and 5:
Call Protec authorized, qualified technical service, if after reset of status message by pressing
the release button (Fig. 2-2/6) the failure status is indicated during next moving again.
2.10.2 Acoustic status message
With all indicator lamp messages (see chapter 2.10.1) an acoustic beeper message is generated.
Beep cycle
Description
2x
General warning
3x
Accumulator nearly
discharged
6x
Accumulator discharged
table 2.3
2.10.3 Acoustic and indicator lamp message if drive blocked
Attention!!
If during moving one drive is blocked (e.g. collision of table top), stop motion and remove
blocked reason
If drive is blocked and reason not found (e.g. internal blocked drive), stop table top elevating
and call Protec authorized, qualified technical service
If one drive is blocked, an optical indicator lamp message and acoustic warning sound take place (see
table 2.2 and 2.3)
If after reset of indicator message with release button (Fig. 2-2/6) the drive blocked reason is not
removed, the drive in function will move only in direction of the blocked drive, independently if up or
down (Fig. 2-2/1, 2 or 7, 8) is commanded. This will prevent inclination of table top.
If the difference between both drives is more than 1cm (0.394inch), moving of drives will be disabled.
The indicator lamp shows cyclic red flashing
and the acoustic beeper warning will become
also cyclic
peep