3.1.9 Exposures with PROGNOST XPE-Akku
For use the operator has to insert a charged accumulator-box (Fig. 2-3) into the power supply box
(Fig. 2-5).
Switch power-supply ON with emergency-stop switch (Fig. 2-1/4).
Out of the controller box comes a short acoustic tone. The green ready for use indicator lamp (Fig. 2-
1/9) must lit.
Attention:
If the green ready for use indicator lamp (Fig. 2-1/9
)
does not lit:
- check correct insertion of accumulator-box into the power supply box
- make sure that emergency-stop switch (Fig. 2-1/4) is switched on
- make sure that accumulator-box is charged
If not possible to solve the problem call Protec authorized service
If the green ready for use indicator lamp (Fig. 2-1/9
)
lit only for approx. 5sec after
power supply is switched ON with emergency-stop switch (Fig. 2-1/4):
- Switch power supply with emergency-stop switch (Fig. 2-1/4) OFF
-
Remove
accumulator-box (Fig. 2-3) out off power supply box (Fig. 2-5
)
call Protec authorized service to check function of electronic inside controller box
Note:
Before removing an discharged accumulator-box, switch power-supply OFF with emergency-
stop switch (Fig. 2-1/4). Out of the controller box comes a short acoustic tone and the green
ready for use indicator lamp do not lit.
The following sequence is necessary in order to avoid a collision between X-ray image receptor and
table top.
•
For exposures with the PROGNOST XPE-Akku bring X-ray unit into desired position.
•
Adjust height of X-ray image receptor to table top height or adjust height of mobile table to
image receptor.
Attention:
The table elevating system is designed for intermittent use only…
•
Move mobile table carefully to desired exposure position and check that there is no collision
between table top, table and image receptor.
•
Move table to position for patient getting ON the table top.
•
Lock brake wheels of mobile table and check that brakes are fixed.
•
Assist patient to get ON the table top.
•
Unlock the brake wheels and move table to previous desired exposure position and lock the
brake wheels.
•
Switch ON collimator light.
•
Move floating table top with patient to final exposure position.
•
Lock brake wheels
•
Collimate the organ and take an exposure.