2.8
Protection foil (optional)
The protection foil on internal covers
(Fig. 2-1/11)
increases the slippage at contact with detector housing.
2.9
Center stop (optional - not shown -)
The center stop for transverse movement of table top improves alignment of the table to x-ray tube assembly.
2.10
Emergency Stop Switch and indicator lamp
By depressing the emergency stop switch (Fig. 2-1/4) the controller and the motor for the elevation of
the table top are switched off.
The emergency stop switch will be released by clockwise rotation.
The three color indicator lamp (Fig. 2-1/9) is placed near the emergency stop switch. It indicates that
the table elevation controller is ready for use and the actual status of the PROGNOST XPE-Akku.
Attention
If the emergency stop switch is not activated and the green ready for use indicator
is off, the electronic inside the table controller box
can be still under voltage!
The table is disconnected from power only by removing the accumulator-box out
of the power supply box.
2.10.1 Status of indicator lamp
indicator lamp low green
XPE energy-saving
mode (see chapter 2.15), depress the
Release switch (Fig. 2-2/6) to activate the XPE to full power
mode.
indicator lamp bright green
XPE ready for operation
indicator lamp orange
Accumulator discharged to first level, normal operation
possible
indicator lamp red
Accumulator discharged, only slow moving possible. After
reaching low power mode it is not possible to switch on to full
power mode back. Change discharged accumulator against
newly charged one.
table 2.1