Protean Instrument Corporation
54
Program MPC
900
Count
Routine
Definition
Page
5
This page will appear only for the following report Models:
•
Standard Report (High Alpha-Only Recount)
•
Flag
Concentration
•
Flag Concentration (High Alpha-Only Recount)
•
Flag Concentration (Radon Subtraction)
This page allows for the ability to set action limits for the alpha only recount and flag
concentration reports.
Action Limits
In environments where many samples are counted within a single shift, it may not be
possible to closely examine the analysis results for every single sample. The MPC 900
can provide a warning on the printed report for samples with excessively high activities.
The alpha action levels may also be used to automatically mark samples for recounting.
The MPC 900 provides separate limits for the alpha and beta channels while in the
“Alpha and Beta” counting mode. If the counting mode is “Alpha Only” only the alpha
action limits are used and the beta action limits are ignored. The action limits are
entered in scientific notation.