Cleaning the machine
Always disconnect the main power from the unit before attempting to clean or repair it.
Dust the lamps with a dry cloth - do not use cleaning fluid or alcohol. The acrylics should be
taken out at least once every week of operation and cleaned thoroughly with a non-alcoholic
disinfectant glass or surface cleaner. If alcohol is used, the acrylic may fade, crack, or break
altogether. Failure to use anything other than an approved cleaning material on the acrylic
will result in poor red light transmission, and could eventually cause injury to the person using
the machine. Use a vacuum cleaner with a dust nozzle to clean around the ballast racks and
throughout the interior portions of the base, top deck, and bench.
Longevity of the acrylic sheets
The acrylic sheets used in this machine should be changed between 1200 and 1500 hours
of operation. If not, transmission could be greatly reduced. ProSun® recommends the use of
Lucasol or Solacryl when cleaning acrylic sheets
Longevity of the lamps
Lamp lifespan is about 1,000 hours.
IMPORTANT
: When replacing or cleaning Facial Lamps, both facial lenses and bulbs should
be cleaned with alcohol, keeping them free of smudges and finger prints
WARNING DANGER
NEVER use this unit if the filter glass is missing, cracked or otherwise damaged. SEVERE
INJURY may result within seconds.
What to do with old lamps?
Contact your local garbage disposal company or city authority to find out what is the most
environmentally sound manner of throwing away old fluorescent lamps.
Care and Maintenance
For more maintenance information please visit our website at: