Page 27
Dulcodes UV-Desinfektionsanlage
Open water drain plug and vent screw and remove
Discharge radiation chamber
Reinstall water drain plug and tighten, only very little force is required for this purpose
Fill the radiation chamber with cleaning solution via the vent opening
Allow the cleaning solution to soak in for at least 20 minutes
Open water drain plug and remove
Discharge (empty) radiation chamber and dispose of cleaning solution according to
requirements
Thoroughly flush radiation chamber with clean water until all remains of the cleaning solution
are removed
Screw in water drain plug and firmly tighten; only very little force is required for this purpose
Slowly open shut-off valve upstream of radiation chamber
Fill radiation chamber until water emerges at the vent screw
Close vent screw; only slight force is required for this purpose
Open shut-off valve downstream of radiation chamber
(necessary only for manual shut-off valve)
Check radiation chamber for leaks
Switch on master switch or connect power plug
The UV disinfection system is now ready for operation.
IMPORTANT!
For radiation chambers that are regularly cleaned by filling with a cleaning solution, it is
recommended to replace the water drain screw and vent screw by suitable valves.
It is advisable to use a suitable acid-resistant pump to fill larger radiation chambers via the
water drain opening.
If the radiation chamber is filled using a pump, it is advisable to circulate the cleaning
solution via the vent opening. This will shorten the cleaning time and improve the results.
7.2
Changing UV lamps
ATTENTION!
The UV lamps should be renewed at the latest when
• the sensor signal approaches the safety threshold provided there are no other causes
such as coating formed on the lamp protective tubes or a substantial deterioration in
UV transmission;
• the operating period of the UV lamps approaches or exceeds the maximum useful life of
the lamps.
Do not touch the glass of the UV lamps with bare hands!
Fingerprints burn into the glass and can cause premature failure. Before fitting the lamps, remove
fingerprints with a cloth moistened with alcohol!
WARNING!
• Switch off the master switch or disconnect the power plug before installing and
connecting the UV lamps!
• UV-C radiation is harmful to the eyes and skin!
Place the UV lamps into operation only in their installed state!
Install the UV disinfection system in accordance with regulations and instructions before
starting up the UV lamps!
• Do not change/modify the lamp connection cables already fitted!
• Do not change/modify the distance between the connector and lamp cover!
Otherwise it will not be possible to ensure that the lamp rests against the closed end of
the protective tube. This position, however, is the prerequisite for safe and reliable
disinfection!
IMPORTANT!
Clean the protective tubes every time the UV lamps are replaced!
For multi-lamp systems: All UV lamps should be replaced in the event of a lamp failure at
the end of the maximum useful life!
Systems without connector on UV lamp cover
Switch off UV disinfection system with the START/STOP button
Switch off master switch or disconnect power plug
Release knurled nut on retaining fixture for UV lamp protective tube by hand
Pull out UV lamp cover until the connector of the connection cable can be disconnected from
the UV lamp
Completely pull out UV lamp and place to one side
Maintenance
BA_DS_009_10_06_GB.p65
20.10.2006, 7:31 Uhr