64
promesstec GmbH
I Niedersachsenstraße 4 I D-48465 Schüttorf I Tel.:+49 (0)5923/ 90 229 0 I Fax:+49 (0)5923/ 90 229 29
E-Mail: [email protected] I Internet: www.promesstec.com
1) Datalogger
Power off, insert the expansion card into the available slot and fix it with an M3
screw.
Insert the prepared microSDHC card in the module after power on. After insertion,
DATALOGGER is created in SET menu of the meter automatically
.
If the meter does not enter the DATALOGGER menu after it is inserted, you can find
this item in the SET menu. Now you can set the activation / deactivation of logging
and the logging interval. Available units for the logging interval are minutes or
seconds.
If the card is accepted and a log file exists at the same time, the end of the file is
searched and the data is added to the end of this file. While the end of the file is
searched, the number of clusters searched and occupied is displayed in the lower left
corner. The number of clusters in a file for searching is limited to 4096 (2MB). If the
log file is longer, a new one is created, increased by 1.
The log file name is LOG00.TXT-LOG99.TXT.
If the data is written correctly, WRITE is displayed for a moment in the lower left corner.
If the microSDHC card is not accepted after logging is activated, an error message is
displayed:
E:1 - GO_IDLE_STATE
E:2 - SEND_IF_COND
E:3 - ACMD41
E:4 - READ OCR
E:5 - no SDHC card
Requirements for microSDHC card:
1)
The card must be SDHC type (cards with 4GB capacity and more)
2)
File format must be FAT32
3)
Cluster size must be 512B*
ON
OFF
min
sec
L
1÷255
L