26) DEFAULT SETTINGS (ORIGINAL FACTORY SETTINGS)
During activation of this function, the configuration of the meter will be restored to
the factory default state in which it was shipped. All user settings will be deleted and
if the metrology jumper J1 on the power supply board is connected (non-certified
meter used for
non-billing
purposes), all volumetric counters will be reset as well.
The user password is cancelled and the access code is reset to original (
0000
).
This applies to calibration of the meter as well. Before activating this function, it is
useful to record or make a back up of the data of all counters.
This function can be activated without the access code!
If you wish to apply the original factory settings, press
E
and use the
or
button to
select YES from menu and then confirm by
E
.
After confirmation of the change, the meter will have the settings it had when it was
delivered by the manufacturer.