5.F.  Kinetic Measurements

The kinetic measurement feature provides a method for monitoring real-time luminescence. The data point frequency 
is the rate at which the measurements are taken for a particular luminescent reaction. Access the kinetic measurement 

options from the Protocol Screen (Figure 5). Touch 

Kinetics

 to turn Kinetics on. Use the up and down arrow keys to 

set the frequency of data point collection as well as the number of total data points. The integration time does not apply 

when the kinetics feature is turned on. Be sure to connect the luminometer to a computer or printer using the RS-232 

serial cable to collect the data points during a kinetic protocol. Touch 

OK

 to save your changes.

5.G.  Default Protocol

Select 

Default Protocol

 to measure your sample for an integration of one second. Touch 

Default Protocol

 to jump 

directly to the Home Screen.

5.H.  Running a User Protocol

To access a saved protocol, select 

Run User Protocol

. The next screen will display protocols previously saved on the 

luminometer. Identify and touch the name of your protocol. The next screen will display the parameters of the saved 

protocol for your review and adjustment. Touch 

OK

 to go to the Home Screen.

 

6. 

Measuring Your Samples

The GloMax

®

 20/20 Luminometer is a bottom-reading luminometer and does not require a minimum sample volume. 

The sample compartment of the luminometer accommodates 35mm petri dishes and 1.5ml microcentrifuge tubes with 
a tube holder. 

1. 

If using a microcentrifuge tube, open the lid of the GloMax

®

 20/20 Luminometer and insert the Microcentrifuge 

Tube Holder. 

2. 

Place your sample in a microcentrifuge tube, and gently place the tube in the microcentrifuge tube holder. Be 
sure that the microcentrifuge tube cap is open. Alternatively, you can place a petri dish directly into the sample 
compartment. 

3. 

Close the lid.

4. 

Optional:

 Touch 

Sample ID

 to name your sample. Using the keypad, enter the sample name into the name field 

(Figure 6). You can enter a maximum of 10 characters. Touch 

Save

 to save the sample ID.

5.  

Touch 

Measure Luminescence

 once to begin the protocol sequence. To abort a measurement in progress, 

touch the screen once.

Summary of Contents for E5311

Page 1: ...Revised 5 21 TM276 T E C H N I C A L M A N U A L GloMax 20 20 Luminometer Instructions for Use of Products E5311 E5351 and E5361...

Page 2: ...ct Components 4 3 Hardware Overview 4 4 Setting up the GloMax 20 20 Luminometer 5 4 A Setup 5 4 B Software Installation and System Connection 5 5 Operating the GloMax 20 20 Luminometer 6 5 A Touch Scr...

Page 3: ...USA 800 356 9526 608 274 4330 Fax 608 277 2516 TM276 Revised 5 21 www promega com 10 Troubleshooting 14 11 Appendix 15 11 A General Care 15 11 B Warranty and Service 15 11 C Specifications 16 11 D Ce...

Page 4: ...horseradish peroxidase and alkaline phosphatase The touch screen on the luminometer allows easy setup and operation The GloMax 20 20 Luminometer also features optional modules for measuring certain fl...

Page 5: ...311 Includes 1 Luminometer 1 1 5ml Microcentrifuge Tube Holder 1 USB Cable 1 Power Supply Kit 1 Quick Start Guide FB081 Items Available Separately P R O D U C T S I Z E C AT GloMax 20 20 Fluorescent M...

Page 6: ...GloMax Spreadsheet Interface Software is compatible with 32 and 64 bit PCs using Windows 7 Professional Service Pack 1 with either Microsoft Office 2010 Home and Business Edition 32 and 64 bit versio...

Page 7: ...a function touch the button corresponding to the function once 5 B Home Screen The Home Screen Figure 3 provides orientation for the multiple functions of the GloMax 20 20 Luminometer From the Home S...

Page 8: ...surement settings automatically programmed Touch OK to accept the parameters and return to the Home Screen You are now ready to measure your samples 5389TA Figure 4 The Protocol Menu 5 E Creating a Ne...

Page 9: ...tocol To access a saved protocol select Run User Protocol The next screen will display protocols previously saved on the luminometer Identify and touch the name of your protocol The next screen will d...

Page 10: ...roll through recent measurements Note Measurements are not saved after power down The instrument display will only show 20 measurements additional measurements will overwrite the first measurements Fo...

Page 11: ...nu Although the touch screen is calibrated at the factory it may need recalibration during instrument setup and over time Contact Promega for more information on screen calibration If the touch screen...

Page 12: ...e Identification Screen will appear Identify the Light Standard The luminometer will automatically power OFF The GloMax 20 20 Luminometer will automatically identify the Light Standard for subsequent...

Page 13: ...ule is designed for easy installation and removal Store the Fluorescent Module in the product case when not in use Keep liquids dust and debris away from direct contact with the Fluorescent Module 9 A...

Page 14: ...ppb ppt or FSU With a valid calibration it is not necessary to generate a standard curve to determine the actual concentration of the sample Calibration Touch Calibration to calibrate the Fluorescent...

Page 15: ...reen until the Home Screen appears Also see Section 8 B Touch Screen Calibration Excel does not open Excel is not installed on the PC Make sure Microsoft Excel is installed on your PC and activated Th...

Page 16: ...11 B Warranty and Service The GloMax 20 20 Luminometer comes with a one year warranty from Promega Additional warranty is available For more information contact Promega Technical Services Contact info...

Page 17: ...ft Office 2007 32 bit version Power 0 5A 100 240V 50 60Hz universal Dimensions 12 92 D 10 44 W 8 42 H 32 82 26 52 21 39cm Weight 8 1 lb 3 65kg Warranty One year CE Compliance Pollution degree 2 Instal...

Page 18: ...on will result in decontamination charges before the instrument will be serviced Select either A or B A I confirm that the returned items have not been contaminated by body fluids or by toxic carcinog...

Page 19: ...20 Replacement Power Supply 1 each E5411 Thermal Serial Printer and Cable Universal Power 1 each E2821 Tripp Lite USB to Serial Adapter 1 each TS2520 Microplate Readers Product Size Cat GloMax Discove...

Page 20: ...rporation Microsoft and Windows are registered trademarks of Microsoft Corporation PicoGreen and RiboGreen are registered trademarks of Molecular Probes Inc Products may be covered by pending or issue...

Reviews: