Notes:
Do not open the instrument lid while a run is in progress.
Touching any part of the screen during measurement will abort the reading.
The GloMax
®
20/20 Luminometer will report the Relative Luminescence Units (RLU) on the Home Screen. If you are
running a Dual-Luciferase
®
Reporter (DLR™) protocol, the ratio for each set of measurements will also appear. The
GloMax™ 20/20 Luminometer will display the status of the last 20 measurements. Use the arrow keys located in the
navigation bar to scroll through recent measurements.
Note:
Measurements are not saved after power down. The
instrument display will only show 20 measurements; additional measurements will overwrite the first measurements.
For more than 20 measurements, connect the instrument to a PC (Section 4.B).
Figure 6. You may name your sample from the Sample ID screen.
7.
Saving Your Protocol
If you wish to store your protocol for future use, touch
Save As
from the Parameters Screen. The Save As screen
contains a keypad to name your protocol. Enter the name of your protocol in the Name field. You can enter a maximum
of 10 characters for your protocol name. Touch
Save
to store the protocol. You can access the saved protocols from the
Run User Protocol feature in Protocols. You may save up to 18 protocols at one time.
5396T
A
Figure 7. You can customize and save your protocol for future use.