6. 

The GloMax

®

 20/20 Luminometer will automatically turn off.

7. 

Turn on the luminometer with the module properly installed. The controls specific to the Fluorescent Module 

will automatically appear on the touch screen.

8. 

Subsequent installation of the Fluorescent Module will automatically launch the proper touch screen controls.

9.B.  Touch Screen Controls

Home Screen: 

From the Home Screen you can access 

Help

Tools

Calibration

Mode

 or 

Measure Fluorescence

Touch the 

Help

 button to access a variety of help topics.

Mode:

 There are two measurement modes for the Fluorescent Module. The Raw Fluorescent mode provides a simple 

method to measure the fluorescence of a sample without calibration. The Direct Fluorescent mode requires a calibration 
with one standard and one blank. The Direct Fluorescent mode reports readings in one of several units (i.e., µg/ml,  
ng/ml, pg/ml, ppb, ppt or FSU). With a valid calibration, it is not necessary to generate a standard curve to determine 
the actual concentration of the sample.

Calibration: 

Touch 

Calibration

 to calibrate the Fluorescent Module with a known standard. Select the units of 

measurement that correspond to the units of your standard. For example, if your standard is 200ng DNA/ml, select the 
ng/ml option. Next, enter the final concentration of your standard. We recommend selecting a standard with a 
concentration similar to a typical sample concentration. Subsequent sample readings reflect the actual concentration.

Measurement: 

Analysis with the Fluorescent Module requires the minicell cuvette. The maximum volume of the 

minicell is 250µl. The minimum volume is 100µl. Insert the minicell cuvette into the Fluorescent Module. Close the lid 
and touch the green 

Measure Fluorescence

 button on the touch screen. After a 5-second measurement time, the 

result will appear on the touch screen next to the Sample ID.

9.C.  Removing the Fluorescent Module

1. 

Turn off the GloMax

®

 20/20 Luminometer.

2. 

Unlock the module by pulling up on the red knob. The module will disengage from the luminometer.

3. 

Carefully remove the Fluorescent Module and store it in the product case

.

Summary of Contents for E5311

Page 1: ...Revised 5 21 TM276 T E C H N I C A L M A N U A L GloMax 20 20 Luminometer Instructions for Use of Products E5311 E5351 and E5361...

Page 2: ...ct Components 4 3 Hardware Overview 4 4 Setting up the GloMax 20 20 Luminometer 5 4 A Setup 5 4 B Software Installation and System Connection 5 5 Operating the GloMax 20 20 Luminometer 6 5 A Touch Scr...

Page 3: ...USA 800 356 9526 608 274 4330 Fax 608 277 2516 TM276 Revised 5 21 www promega com 10 Troubleshooting 14 11 Appendix 15 11 A General Care 15 11 B Warranty and Service 15 11 C Specifications 16 11 D Ce...

Page 4: ...horseradish peroxidase and alkaline phosphatase The touch screen on the luminometer allows easy setup and operation The GloMax 20 20 Luminometer also features optional modules for measuring certain fl...

Page 5: ...311 Includes 1 Luminometer 1 1 5ml Microcentrifuge Tube Holder 1 USB Cable 1 Power Supply Kit 1 Quick Start Guide FB081 Items Available Separately P R O D U C T S I Z E C AT GloMax 20 20 Fluorescent M...

Page 6: ...GloMax Spreadsheet Interface Software is compatible with 32 and 64 bit PCs using Windows 7 Professional Service Pack 1 with either Microsoft Office 2010 Home and Business Edition 32 and 64 bit versio...

Page 7: ...a function touch the button corresponding to the function once 5 B Home Screen The Home Screen Figure 3 provides orientation for the multiple functions of the GloMax 20 20 Luminometer From the Home S...

Page 8: ...surement settings automatically programmed Touch OK to accept the parameters and return to the Home Screen You are now ready to measure your samples 5389TA Figure 4 The Protocol Menu 5 E Creating a Ne...

Page 9: ...tocol To access a saved protocol select Run User Protocol The next screen will display protocols previously saved on the luminometer Identify and touch the name of your protocol The next screen will d...

Page 10: ...roll through recent measurements Note Measurements are not saved after power down The instrument display will only show 20 measurements additional measurements will overwrite the first measurements Fo...

Page 11: ...nu Although the touch screen is calibrated at the factory it may need recalibration during instrument setup and over time Contact Promega for more information on screen calibration If the touch screen...

Page 12: ...e Identification Screen will appear Identify the Light Standard The luminometer will automatically power OFF The GloMax 20 20 Luminometer will automatically identify the Light Standard for subsequent...

Page 13: ...ule is designed for easy installation and removal Store the Fluorescent Module in the product case when not in use Keep liquids dust and debris away from direct contact with the Fluorescent Module 9 A...

Page 14: ...ppb ppt or FSU With a valid calibration it is not necessary to generate a standard curve to determine the actual concentration of the sample Calibration Touch Calibration to calibrate the Fluorescent...

Page 15: ...reen until the Home Screen appears Also see Section 8 B Touch Screen Calibration Excel does not open Excel is not installed on the PC Make sure Microsoft Excel is installed on your PC and activated Th...

Page 16: ...11 B Warranty and Service The GloMax 20 20 Luminometer comes with a one year warranty from Promega Additional warranty is available For more information contact Promega Technical Services Contact info...

Page 17: ...ft Office 2007 32 bit version Power 0 5A 100 240V 50 60Hz universal Dimensions 12 92 D 10 44 W 8 42 H 32 82 26 52 21 39cm Weight 8 1 lb 3 65kg Warranty One year CE Compliance Pollution degree 2 Instal...

Page 18: ...on will result in decontamination charges before the instrument will be serviced Select either A or B A I confirm that the returned items have not been contaminated by body fluids or by toxic carcinog...

Page 19: ...20 Replacement Power Supply 1 each E5411 Thermal Serial Printer and Cable Universal Power 1 each E2821 Tripp Lite USB to Serial Adapter 1 each TS2520 Microplate Readers Product Size Cat GloMax Discove...

Page 20: ...rporation Microsoft and Windows are registered trademarks of Microsoft Corporation PicoGreen and RiboGreen are registered trademarks of Molecular Probes Inc Products may be covered by pending or issue...

Reviews: