4. 

Setting up the GloMax

®

 20/20 Luminometer

4.A. Setup

1. 

Place the GloMax

®

 20/20 Luminometer on a flat, level surface. Allow at least 6 inches (16cm) of vertical clearance 

to open and close the lid. Position the luminometer so that the touch screen faces you.

2. 

Connect the power supply into the back of the luminometer. 

3. 

Plug the power cord into a wall outlet. See Section 11.C for power requirements.

4. 

Turn the On/Off switch to the “On” position (Figure 2).

 

Note: Touch screen calibration can drift during shipment. If this happens, touch screen alignment will be affected 
and buttons may be difficult to press or unresponsive. See Section 8.B for Touch Screen Calibration.

4.B.  Software Installation and System Connection

To export data from the luminometer to an Excel spreadsheet, you must first install the GloMax

®

 Spreadsheet Interface 

Software on a PC with an available serial port or USB connector. The GloMax

®

 Spreadsheet Interface Software is 

compatible with 32-and 64-bit PCs using Windows

®

 7 Professional Service Pack 1 with either Microsoft

®

 Office 2010 

Home and Business Edition (32- and 64-bit versions) or Microsoft

®

 Office 2007 (32-bit version). 

1. 

Download the software from the Promega website: 

www.promega.com/resources/software-firmware/

glomax-systems/glomax-20-20-luminometer/

. This will initiate the installation program. When  

installation is complete, the GloMax

®

 Spreadsheet Interface Software icon will appear on your PC desktop and in 

the Programs menu.

2. 

To establish a connection between the luminometer and the PC, connect the USB cable between the luminometer 
and your PC. If connecting to the PC with a USB/RS-232 adaptor, we recommend using the Tripp-Light, Model# 
U209-000-R, which can be purchased from Tripp Lite (Cat.# U209-000-R), or from Promega (Cat.# TS2520). 
When using a USB-Serial adaptor, you may need to download the appropriate computer driver available at:  

www.tripplite.com

. Select the PC COM port in the software using the Select COM port button to establish a 

connection with the instrument.

 

Alternatively, connect the luminometer to a printer to print the data as they appear onscreen. A thermal printer 
(Cat.# E2821) is available separately from Promega.

Summary of Contents for E5311

Page 1: ...Revised 5 21 TM276 T E C H N I C A L M A N U A L GloMax 20 20 Luminometer Instructions for Use of Products E5311 E5351 and E5361...

Page 2: ...ct Components 4 3 Hardware Overview 4 4 Setting up the GloMax 20 20 Luminometer 5 4 A Setup 5 4 B Software Installation and System Connection 5 5 Operating the GloMax 20 20 Luminometer 6 5 A Touch Scr...

Page 3: ...USA 800 356 9526 608 274 4330 Fax 608 277 2516 TM276 Revised 5 21 www promega com 10 Troubleshooting 14 11 Appendix 15 11 A General Care 15 11 B Warranty and Service 15 11 C Specifications 16 11 D Ce...

Page 4: ...horseradish peroxidase and alkaline phosphatase The touch screen on the luminometer allows easy setup and operation The GloMax 20 20 Luminometer also features optional modules for measuring certain fl...

Page 5: ...311 Includes 1 Luminometer 1 1 5ml Microcentrifuge Tube Holder 1 USB Cable 1 Power Supply Kit 1 Quick Start Guide FB081 Items Available Separately P R O D U C T S I Z E C AT GloMax 20 20 Fluorescent M...

Page 6: ...GloMax Spreadsheet Interface Software is compatible with 32 and 64 bit PCs using Windows 7 Professional Service Pack 1 with either Microsoft Office 2010 Home and Business Edition 32 and 64 bit versio...

Page 7: ...a function touch the button corresponding to the function once 5 B Home Screen The Home Screen Figure 3 provides orientation for the multiple functions of the GloMax 20 20 Luminometer From the Home S...

Page 8: ...surement settings automatically programmed Touch OK to accept the parameters and return to the Home Screen You are now ready to measure your samples 5389TA Figure 4 The Protocol Menu 5 E Creating a Ne...

Page 9: ...tocol To access a saved protocol select Run User Protocol The next screen will display protocols previously saved on the luminometer Identify and touch the name of your protocol The next screen will d...

Page 10: ...roll through recent measurements Note Measurements are not saved after power down The instrument display will only show 20 measurements additional measurements will overwrite the first measurements Fo...

Page 11: ...nu Although the touch screen is calibrated at the factory it may need recalibration during instrument setup and over time Contact Promega for more information on screen calibration If the touch screen...

Page 12: ...e Identification Screen will appear Identify the Light Standard The luminometer will automatically power OFF The GloMax 20 20 Luminometer will automatically identify the Light Standard for subsequent...

Page 13: ...ule is designed for easy installation and removal Store the Fluorescent Module in the product case when not in use Keep liquids dust and debris away from direct contact with the Fluorescent Module 9 A...

Page 14: ...ppb ppt or FSU With a valid calibration it is not necessary to generate a standard curve to determine the actual concentration of the sample Calibration Touch Calibration to calibrate the Fluorescent...

Page 15: ...reen until the Home Screen appears Also see Section 8 B Touch Screen Calibration Excel does not open Excel is not installed on the PC Make sure Microsoft Excel is installed on your PC and activated Th...

Page 16: ...11 B Warranty and Service The GloMax 20 20 Luminometer comes with a one year warranty from Promega Additional warranty is available For more information contact Promega Technical Services Contact info...

Page 17: ...ft Office 2007 32 bit version Power 0 5A 100 240V 50 60Hz universal Dimensions 12 92 D 10 44 W 8 42 H 32 82 26 52 21 39cm Weight 8 1 lb 3 65kg Warranty One year CE Compliance Pollution degree 2 Instal...

Page 18: ...on will result in decontamination charges before the instrument will be serviced Select either A or B A I confirm that the returned items have not been contaminated by body fluids or by toxic carcinog...

Page 19: ...20 Replacement Power Supply 1 each E5411 Thermal Serial Printer and Cable Universal Power 1 each E2821 Tripp Lite USB to Serial Adapter 1 each TS2520 Microplate Readers Product Size Cat GloMax Discove...

Page 20: ...rporation Microsoft and Windows are registered trademarks of Microsoft Corporation PicoGreen and RiboGreen are registered trademarks of Molecular Probes Inc Products may be covered by pending or issue...

Reviews: