For example, Measure a batch of CMOS voltage data signal in channel of
CH00~CH03 and a batch of 1V voltage data in CH04~CH07. Threshold voltage
setting steps as below:
Press "
1 (threshold)
" and menu appears.
Press "
F1
" till channel source display as "
CH00~CH03
".
Press "
F2
" to choose threshold as "
CMOS/(2.5V)
".
Press "
F1
" till channel source display as "
CH04~CH07
".
Press "
F2
" to choose threshold as "custom". Menu setting correspond
to
F3
display the setting voltage value.
Turn "
CH1 Volts/div
" knob and set voltage value as "
0.50V
".
Threshold voltage setting finished (refer to Fig. 88).
Figure 88.
Threshold setting menu.