Page
64
October 2014
3.5.1
How to Implement the Auxiliary System Function Setting
Press the
UNTILITY
button and the menu is displayed in the screen as Fig. 60.
Figure 60.
Function menu.
Function Menu
Setting
Description
System Status
Display the system function menu.
Recall Factory
Call out the factory settings.
Do Self Cal
Carry out the self-calibration procedure.
Language
Spanish
English
…
Choose the display language of the
operating system.
►
Carry out a Self-Calibration
We have a built-in Self-Calibration procedure, by performing a Self-Calibration
would help to maintain the accuracy of your oscilloscope under ambient
temperature. It is recommended to carry a self-Calibration if the change of the
ambient temperature is up to or exceeds 5
°
C.
NOTE: Disconnect all inputs before carrying out a self-calibration. Fail
to do so may cause damage to your oscilloscope.