Slope Trigger menu list:
Menu
Settings
Instruction
Source
CH1
CH2
Channel 1 as the trigger source.
Channel 2 as the trigger source.
Mode
Slope
when
Set slope condition.
Settings
24 ns
∼
10 s
Turn "TRIG LEVEL" knob to set slope time.
High level
Turn "TRIG LEVEL" knob to set the High level.
Low level
Turn "TRIG LEVEL" knob to set Low level.
Slew rate
Slew rate=( High level- Low level)/ Settings.
Trigger
mode
Auto
Normal
Single
Acquire waveform whatever detect trigger
condition or not.
Only acquire waveform when match trigger
condition.
Only acquire waveform for single time when detect
trigger condition then stop.
Holdoff
100 ns
∼
10 s
Set interval by using TRIG LEVEL control, value
range from 100ns
∼
10s.
Holdoff
Reset
Reset hold time to 100ns