October 2014
Page
11
2.2.4
How to Implement the Probe Compensation
When connect the probe with any input channel for the first time, make this
adjustment to match the probe with the input channel. The probe which is not
compensated or presents a compensation deviation will result in the measuring
error or mistake. For adjusting the probe compensation, please carry out the
following steps:
Set the attenuation coefficient of the probe in the menu as 10X and that
of the switch in the probe as 10X, and connect the oscilloscope probe
with the CH1 channel. If a probe hook tip is used, ensure that it keeps in
close touch with the probe. Connect the probe tip with the signal
connector of the probe compensator and connect the reference wire
clamp with the ground wire connector of the probe connector, and then
press the button "7(AUTOSET)".
Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Figure 5 and Figure 6).
Figure 6
. Displayed Waveforms of the Probe Compensation.
Repeat the steps mentioned if needed.
Figure 7.
Adjust Probe.