4016 Series Operation Manual 47
COMMAND
RETURN
備註
OUT?{; | NL}
ON | OFF
0=OFF,1=ON
MODE?{; | NL}
AC | DC
METER?{; | NL}
0~7
0=Menu,1=Meter,2=Harmonic,3=Inrush
,4=AC Standby,5=DC Accumuloator
6=Data
Log
,7=On/Off Cycling
VRANG?{; | NL}
1~6
1=20V,2=40V,3=80V,4=200V,
5=400V,6=800V
IRANG?{; | NL}
1~18
1=2mA,2=4mA,3=8mA,4=20mA,
5=40mA,6=80mA,7=0.2A,8=0.4A,
9=0.8A,10=2A,11=4A,12=8A,13=10A
,14=20A,15=40A,16=50A,17=100A,
18=200A
SHUNT?{; | NL}
INT | EXT
FILTER?{; | NL}
ON | OFF
ONDEG?{; | NL}
0~359
OFFDEG?{; | NL}
0~359
GraphT?{; | NL}
0~100ms
VERsion?{; | NL}
r#.##,r#,r#
Display rev. , Module1 rev. , Module rev.
*IDN?{; | NL}
PRODIGIT:4016
ONTIME?{; | NL}
###.### (sec)
OFFTIME?{; | NL}
###
.### (sec)
REPEAT?{; | NL}
####
SCALE?{; | NL}
1~10000
AUTOUP?{; | NL}
ON | OFF
THD?{; | NL}
0 | 1
0=THDR ,1=THDF
GRAPH?
0 | 1
0=AVG,1=OR