4016 Series Operation Manual 51
KWH?
Syntax:
MEAS: KWH? {; | NL}
Description
:
Read accumulated watt hour value.
AVGWATT?
Syntax:
MEAS: AVGWATT? {; | NL}
Description
:
Reading watt average.
ELT?
Syntax:
MEAS: ELT? {; | NL}
Description
:
Read elapsed time.
INRUSHV?
Syntax:
MEAS: INRUSHV? {; | NL}
Description: Read and measure the Inrush voltage value.
INRUSHI?
Syntax:
MEAS: INRUSHI? {; | NL}
Description: Read and measure the Inrush current value.
GROUP?
Syntax:
MEAS: GROUP? {; | NL}
Description: Read measured values (Vrms, Vpk+, Vpk-, Vmax, Vmin, Irms, Ipk+, Ipk-,
Imax, Imin, Watt, Wmax, Wmin, VA, VAR, PF, VCF, ICF, Hz).
GRAPH?
(For RS232 or USB Interface only)
Syntax:
MEAS: GRAPH? {; | NL}
Description: 3bytes x4096(Volt Graph)+3bytes x4096(Current Graph)+5bytes x
4096(Watt Graph)+0x0D 0x0A
Graphic data format: the 1st highest bit is positive and negative flag, 0=positive value
1=negative value, the value contains decimal place, the decimal place is determined
according to the current gear position (please refer to 1.6 Specifications), the watt
decimal place is composed of voltage and current gear Decide
Volt Graph (VRange=400V):0x00 0x2A 0xF8=110.00V
Current Graph(IRange=10A):0x80 0x1F 0x40= -8.000A
Watt Graph(400V 10A
檔位
):0x80 0x05 0x3E 0xC6 0x00=-880.00000W
※
Before reading , you need to send command
〝
LOCK ON
〞
to lock the data update.
At this time, the measurement data will not be updated. After reading , you need to send
command
〝
LOCK OFF
〞
to unlock the data.
The starting point of the graphic data is random, not starting from 0 degrees
V GRAPH?
(For RS232 or USB Interface only)
Syntax: MEAS: VGRAPH?{; | NL}
Description: 3bytes x4096(Volt Graph)+0x0D 0x0A
The 1st highest bit is positive and negative flag, 0=positive value 1=negative value, the
value contains decimal place, and the decimal place is determined according to the
current gear position (please refer to 1.6 Specifications)
Volt Graph (VRange=400V):0x00 0x2A 0xF8=110.00V
※
Before reading , you need to send command
〝
LOCK ON
〞
to lock the data update.
At this time, the measurement data will not be updated. After reading , you need to send
command
〝
LOCK OFF
〞
to unlock the data.