METIS Vision MV05
/ MV09 Rugged Thermovision Camera
21
The colour of the cable cores specified in the tables
below corresponds to the pin assignment when using
the original accessory cables. If other cables are used
the colour code can deviate.
The designation of the connector pins is based on the
pin assignment on the device.
Connecting the supply voltage
Personnel:
Trained electrician
NOTICE!
Danger of damage to the Thermovision Camera due
to the supply voltage!
The Thermovision Camera can be damaged if the supply
voltage is incorrectly connected or if the connection is
defective.
Switch off the supply voltage prior to starting the
tasks.
Comply
with the supply voltage specification
(
Chapter 3.3
‘Connected loads’ on page 10).
Supply Voltage hast o be fused by 800 mA slow
blow.
Ensure the correct pin assignment for the connection.
Proceed as follows to establish the connection to the supply voltage
Establish the connections of the Thermovision Camera to the
connection of the supply voltage as specified in the table below:
Fig. 6: Connector pins on the plug
connector of the Thermovision
Camera
Connector
pin
Colour of cable
core
Supply voltage connections
K
White
24V supply voltage (+)
A
Brown
0V supply voltage (-)
Power can also be supplied via an external power adapter
that meets the specifications listed in the technical data
(
Chapter 3.3
‘Connected loads’ on page 10
).
Connecting the trigger input
Personnel:
Trained electrician
The Trigger input of the camera is galvanically separated from the
operating voltage by an optocoupler. Pin A (brown) and Pin G (red) of
the plug have to be connected, the power supply voltage must be
conneceted to Pin J (pink).