Prism
Operation Manual Issue 2.00
Page 6.45
The [EXIT] key terminates the logging operation and leaves Watchdog Run to Log
mode. Note that an '<END OF TEST>' event is placed in the log at this point in order
that the duration of the test is recorded.
After running the Watchdog, the format of the Results Log is:
Prism Sound DSA-1 Results Log:
------------------------------
*** BLOCK 01 ***:
[Watchdog:]
001. 001:12:10.0 C-S:CRC Error
002. 001:12:13.4 C-S:CRC OK
003. 002:20:28.8 C-S:Consumer
004. 003:45:48.9 Main Aud Black
005. 003:45:51.1 Main Aud Activ
006. 008:43:55.6 <END OF TEST>
It is possible to set up the DSA-1 so that Watchdog Events are accompanied by the
transmission of an Event-specific serial code from the RS-232 port. This is done by
enabling 'Alarms' in the System Settings sub-menu as described in section 6.9.1.4. If
Alarms and Key Beeps are both enabled, each Event is also accompanied by a beep
from the DSA-1's loudspeaker.
NOTE: Some LEDs only cause an Event to be logged when they come on , and not
when they go off (these are the 'ERROR', 'N.FAIL', 'BLOCK', 'PRO' and 'CONS'
LEDs); all the others cause Events to be logged both as they are lit and as they are
extinguished. A special case is the 'FAIL' LED which always causes an Event as it
lights whether it is alarmed or not, but only causes an Event when it goes out if it is
alarmed.
Refer to appendix A for a list of possible Events in Watchdog mode, and their Alarm
codes.