16.
Recommended Cleaning Instruction
General cleaning
The exterior of the Hoist can be cleaned using a damp soapy cloth for general
cleaning duties.
Please ensure the cloth is damp and not wet.
Ensure the
exterior of the device is dry after cleaning, dry using a clean dry cloth.
Care should always be taken when cleaning around electrical components
Disinfecting (if necessary)
Should the Hoist require a more thorough clean, the use of the Actichlor
disinfectant product, which is widely available in tablet form and used
throughout the healthcare industry, is recommended for use to ensure a
thorough clean.
•
Please follow the manufacturers safety instruction for the use of the
cleaning product before use to ensure safe use for the operator and the
patient.
•
Ensure the cloth is
damp
before the cleaning process.
DO NOT
use
a wet cloth over electrical systems.
•
Be careful not to let water ingress into the device as although the
device is IP rated, it is not water tight.
Application is through a clean soaked (but damp) cloth applied to
wipe the device down
Used in the following dilutions to ensure an effective clean:
•
Actichlor dissolvable chlorine tablets provide a concentration of
1000 ppm of available chlorine (0.1%) per 1 tablet
•
1 tablet (1.7g formed tablet (x1)) will create a virucidal solution,
diluted in 1 litre of water to provide effective means to clean a
“Dirty” device. This is also ideal for use after an outbreak of the
Norovirus / winter vomiting and can be used as a precaution
against C.Diff. It is effective against viruses, bacteria, spores,
yeasts and moulds.
•
The contact time against the outer components of the device
should be for 5 minutes to prevent any virucidal infections without
a degradation to the functionality of the device. 5 minutes is a
recommended contact time. The device can withstand a longer
contact period but the 5 minute recommendation as a minimum
must be followed to provide an effective cleaning regime.
•
Blood spills should be dealt with by an increased concentration of
the solution – please refer to the instructions on the
manufacturers product labelling.
Summary of Contents for Transactive Xtra
Page 1: ...Transactive Xtra User Manual...
Page 35: ...Transactive Xtra User Manual Page 35 of 40 Part No 999015 Rev 2 Nov 2017...
Page 36: ...Transactive Xtra User Manual Page 36 of 40 Part No 999015 Rev 2 Nov 2017...
Page 37: ...Transactive Xtra User Manual Page 37 of 40 Part No 999015 Rev 2 Nov 2017...
Page 38: ...Transactive Xtra User Manual Page 38 of 40 Part No 999015 Rev 2 Nov 2017...
Page 39: ...Transactive Xtra User Manual Page 39 of 40 Part No 999015 Rev 2 Nov 2017 User notes...