18.6.2
Assemble the detector
Only use isolating plastic screws to fasten the detector to prevent any electric noise signals from
being interspersed. Do not forget the foam rubber spacer during installation, otherwise the rota-
tional disc could be mechanically blocked by the screws. The foam rubber spacer also ensures
mechanical decoupling of the detector.
1. To install the new detector, first place the foam rubber spacer (C) on the mounting surface of the detec-
tor. Then connect the cables. Align the detector in such a way that a space is maintained to the housing
on both sides (see Fig. 18.8 on page 56).
NOTICE
Blocking the rotational disc
If the screws are tightened too firmly, they might block the rotational disc!
X
Only tighten the screws hand-tight. The foam rubber spacer must only be compressed to no
more than half its original thickness.
2. Fasten the detector with the two plastic screws (D).
Maintain distances!
C
D
Fig. 18.8:
Assembling a new detector