![Primes FocusMonitor FM+ Operating Manual Download Page 20](http://html1.mh-extra.com/html/primes/focusmonitor-fm/focusmonitor-fm_operating-manual_1601726020.webp)
FocusMonitor FM+
35,0(6
20
Revision 03/2019 EN
7.5
Install the FocusMonitor FM+
DANGER
Serious eye or skin injury due to laser radiation
If the device is moved from its calibrated position, increased scattered or directed reflection
of the laser beam occurs during measuring operation (laser class 4).
X
Mount the device so that it cannot be moved by an unintentional knock or cables being
pulled accidentally.
The mounting surface of the housing has six slotted holes Ø 6.4 mm and four alignment holes Ø 6
G7
mm for
assembly on a support bracket provided by the customer (see Fig. 7.9 on page 20).
Use at least four M6 screws to fasten the housing.
The total length of the screws depends on the dimensions of the customer’s support bracket.
The dimensioned arrangement of the mounting holes is specified in chapter 17, „Dimensions“, on
page 44.
Slotted holes Ø 6.4 mm
Alignment holes Ø 6
G7
mm
Fig. 7.9:
Mounting holes, view from above (same hole pattern below)
NOTICE
Damage/Destruction of the measuring tip
If the laser beam hits the measuring tip in the beam path, it could be destroyed.
X
Remove the measuring tip from the beam path following installation.