![Primes FocusMonitor FM+ Operating Manual Download Page 24](http://html1.mh-extra.com/html/primes/focusmonitor-fm/focusmonitor-fm_operating-manual_1601726024.webp)
FocusMonitor FM+
35,0(6
24
Revision 03/2019 EN
8.4
Connection of the FocusMonitor FM+ and PowerMonitor PM 48/100 to the PC
For full absorption of the radiation behind the measurement zone, you can use the PRIMES PowerMonitor
PM 48/100. The water-cooled PowerMonitor PM 48/100 will show you the power, water flow rate, and water
temperature.
Ethernet
PC
FocusMonitor FM+
PRIMES power supply
PowerMonitor PM 48/100
PRIMES-Bus
RS485
L
Max
=10 m
Fig. 8.3:
Connection of the FocusMonitor FM+ and PowerMonitor PM 48/100 to the PC
NOTICE
Damage/Destruction of the device due to overvoltage
When disconnecting the electric cables during operation (when the supply voltage is con-
nected), voltage peaks can be generated that could destroy the communication modules of
the measuring devices.
X
Turn off the power supply before disconnecting the bus cables.
Only ever use one power supply (typically original PRIMES power supply) when connecting several
devices.