List of error codes
METRAX GmbH
DefiMonitor XD 21324 / GB / K05
83 / 116
87
Program update error
Error during program update, the
update will be interrupted
- return the device for servicing
90
Tasks start/run error
Error during internal processing
- restart the device
- if the error persists: Return the
device for servicing
91
Menu init error
Defective display in the Setup
menu
- return the device for servicing
92
Critical key error
Error during the check of input
elements (e.g. shock button)
- restart the device
- if the error persists: Return the
device for servicing
93
Charge time-out error
The high voltage charge lasts
too long, possibly due to a
failure in the final stage
- return the device for servicing
94
Disarm time-out error
Impossible to discharge the
residual energy in the
condensator, possibly due to a
failure in the final stage
- return the device for servicing
96
AVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
97
AVDD drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
98
CVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
99
CVDD drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
100
DVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
101
DVDD low
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
102
5V exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing
103
5V drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the