Rhythm detection system in AED mode
METRAX GmbH
DefiMonitor XD 21324 / GB / K05
103 / 116
Sensitivity
= Number of "correct
shockable
" algorithm decisions
Total number of ECGs in which dispensing an impulse is clinically recommended
Specificity
= Number of "correct
not shockable
" algorithm decisions
Total number of ECGs where shock is not clinically recommended
False positive rate
= Number of "incorrect
shockable
" algorithm decisions
Total number of ECGs where shock is not clinically recommended
Positive predictive value
= Number of "correct
shockable
" algorithm decisions
Total number of ECGs in which administering an impulse is recommended by the device