PRESYS
Instruments
T-650PH
Page 10
TU
: Menu that configures the PID control parameters for the heating side. More
details in section
6 - PID Control Parameters
.
K
I
D
FACT
K
: Proportional gain
I
: Integral gain
D
: Derivative gain
FACT
: Restore the control parameters to the factory values.
PRG
: Menu that programs the calibrator.
DEC_IN
DEC_PRB
SP
PRBe
MD
DEC_IN
: Selects the number of decimals of the thermo-element reading.
The default value depends on the input signal.
DEFAULT
0
1
2
3
4
DEC_PRB
: Selects the number of decimals of the block temperature and
the setpoint value. The default number is 2.
0
1
2
SETPOINT
: Enables the calibrator
Programmable Operating Mode
, and
allows the configuration of programmed values. The selected program is
indicated by the selection arrow. Choose one of the
6
temperature programs or
NO
to disable this
Programmable Mode
.
Select any of the 6 programs and confirm with the key
ENTER.
Following, it is
shown the menu of temperature setpoint configuration.
10%
20%
25%
VARIABLE
Change the configuration to steps (STEPS) of
10%
,
20%
,
25%
,
VARIABLE
or
press
C/CE
to maintain the configuration already stored in memory. The temperature
range of the program must be configured through the values in
SETPOINT HIGH
and
SETPOINT LOW
in case of
10%
,
20%
or
25%
fixed steps of the range. The option
VARIABLE
allows the user to define from 2 to 11 values of temperature setpoint, not
necessarily in ascending order.
presys