Page 17
3.2 - Recommendations to Other types of Insert
When the user uses a different type of insert, other than the one used to adjust
the dry block, the accuracy from the manual is no longer valid, since the thermal contact
is different in relation to that insert.
However, another control sensor adjustment can be performed to the new insert.
Do that using an external probe with high accuracy, placed in the new insert and follow
the steps described in item
7.2 - Probe Calibration
.
This new readjustment of the calibrator indication overwrites the factory calibration
and it remains even after turning off the dry block.
Should you want the factory calibration back again, activate the option
RESTORE
from menu calibration / probe.