PRESYS
Instruments
T-650PH
Page 2
T-650PH has also many other features, such as:
The electric signal calibrator is independent from the dry block function.
Internal buzzer beeps when the temperature reaches the desired value.
Keypad that eases the operation and configuration of the calibrator.
Graphic Fluorescent vacuum display to present big-size digit numbers.
Thermo-element reading scaled to ITS-90 or IPTS-68.
Internal regulated 24 Vdc power supply for 2-wire transmitters.
Internal rechargeable battery and battery charger included in the electric signal
calibrator.
Independent circuitry for over-temperature protection and safety.
Insert to choose, carrying case, strap and test leads included.
1.1 - Technical Specifications
Operating Range
from room temperature to 650 °C (1202 °F)
Power Supply
110Vac or 220Vac 50/60Hz,
according to order code.
Well Depth
124 mm
Resolution
0.01 °C (0.01 °F)
Accuracy
I
nternal reference
with
external probe
with
external thermometer
0.1 °C
0.1 % of reading
0.10 °C
0.05 °C
Stability
(after 10 minutes)
0.05 °C
Power Consumption
1000 W
Heating Rate
20 minutes
(50 °C
to
650 °C)
Cooling Rate
21 minutes
(650 °C
to
200 °C)
Dimensions (H,W,D)
215 x 390 x 310 mm
Weight
9.0 kg
Warranty
1 year
presys