PremiumLift AirCare Max-Pro ACMaxPP User Manual Download Page 8

8

 

www.premiumlift.com.au 

 

 
 
 
 
 
 
 
 

The mains cable must be 

free and not be allowed to be caught up in the bed’s 

moving mechanisms. The mains cable may be damaged or it may be pulled out of 
its socket and electric leads exposed as a result.

 

 

If the mains cable or the mains plug are damaged, the relevant part must be 
replaced. This work should be carried out by the manufacturer or authorised 
service agents.

 

 

When connecting the mains plug do not use multiple sockets since liquids may

 

penetrate into these (fire hazard and electric shock).

 

 
 

 

3.2 

SAFETY INFORMATION FOR THE  OPERATOR 

 

With the help of this instruction manual, instruct each user in the safe operation of

 

this system before it is put into service for the first time.

 

 

Advise the user of any hazards which may occur if not handled correctly.

 

 

Only persons who have been properly instructed may operate this system. This 
also applies for persons who only operate the system on a temporary basis.

 

 
 

 

3.3 

SAFETY INFORMATION FOR THE  USER 

 

Ensure that the operator instructs you in the safe operation of this system.

 

 

In addition, pay particular attention to the ‘Warnings and Cautions’ (

page 4-5

) and the 

‘General Safety Information’ as described in 

3.1

.

 

 

If there is a suspected fault or damage, unplug the mains plug from the socket and follow 
the power down procedure (

page 16

). Clearly mark ‘Out of Order’, take out of service 

immediately and inform the person in charge without delay.

 

GENERAL 

INFORMATION

 

Summary of Contents for AirCare Max-Pro ACMaxPP

Page 1: ...AirCare Max Pro Digital Pump User Manual...

Page 2: ...2www premiumlift com au...

Page 3: ...AFETY INSTRUCTIONS 3 2 SAFETY INFORMATION FOR THE OPERATOR 3 3 SAFETY INFORMATION FOR THE USER 3 4 SYMBOLS USED 3 5 CLEANING DISINFECTION 3 6 SERVICING MAINTENANCE 3 7 SERVICE LIFE DISPOSAL TECHNICAL...

Page 4: ...the therapy may be reduced All hoses must be free of kinks twists and must be properly connected and positioned so as not to cause any obstruction Do not position the pump in a way that prevents acce...

Page 5: ...found on page 15 Suitableforcontinuous use Do not modify the mattress cushion or pump unit in any way Do not connect to any other medical device or equipment Not for use in an oxygen enriched environ...

Page 6: ...g pressure surface designed to reduce surface pressure on the skin and therefore reduce risk of the development of pressure ulcers in susceptable patients This pump has audible and visual alarms for l...

Page 7: ...ltoftheirvocational training or briefing are entitled to deliver assemble disassemble and transport the product 2 NON COMPLIANT USE All uses deviating from the intended purpose which may also be hazar...

Page 8: ...ct each user in the safe operation of this system before it is put into service for the first time Advise the user of any hazards which may occur if not handled correctly Only persons who have been pr...

Page 9: ...ccording to EN 60601 Handle with care This way up Keep dry Recycling symbol Refers to packaging that can be recycled cardboard Fragile handle with care This product must be disposed of in a designated...

Page 10: ...cts damageorsignsofwearmustberectifiedwithoutdelay Onlyoriginal spare parts from PremiumLift may be used otherwise all guarantees or warranties will be invalidated See pages 16 17 for Routine Maintena...

Page 11: ...6 9 12 25 minutes 345 x 115 x 185mm 3 5 kg 14 lpm Manual PUMP POWER REQUIREMENTS Power Rating Voltage Fuse MedicalClassification Safety Standards 13W AC 240V 50Hz T 1AH 250V Medical Device Class 1 EN...

Page 12: ...Pipes from the mattress connect here The pump must be able to be disconnected in an emergency for CPR to be administered B Power Switch The power switch is on the left hand side of the pump When press...

Page 13: ...ansfers Press the Cycle Time button until all 4 LED will illuminate green for Static Mode H Maxi Inflate Mode Press the Maxi Inflate button to automatically inflate the mattress to its maximum pressur...

Page 14: ...ug from the mains supply TRANSPORT MODE If the pump needs to be disconnected for any reason attach the transport cap to the feed tubes connector This will prevent deflation retaining the remaining air...

Page 15: ...s mattressesshouldbethoroughlyrinsedwithcleanwatertoremoveany residues This will help prevent any long term compatibility issues associated with disinfectant residues Alternatively disinfectionmay bea...

Page 16: ...ed clean with a mild antiseptic solution All cells are replaceable and can besourced from PremiumLift POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance r...

Page 17: ...are Max Pro pump should be serviced every 12 months by PremiumLift approved personnel using genuine PremiumLift spare parts Failure to follow the service schedule may invalidate future warranty claims...

Page 18: ...e mains plug is not blown 4 The wall socket that the pump is connected to is working correctly 1 Connect the pump to the nearest working mainsoutlet 2 Replace the plug fuse with the correct fuse 3 Try...

Page 19: ...is not trapped in the bed mechanism Contact PremiumLift technical support Before calling Pleaseensureyouhavetheserial number and model of equipment 2 The CPR valve is not trapped and is in the closed...

Page 20: ...not appear to be alternating 1 Check that the static mode is not turned on 2 Carefully mark one ofthe inflated cells with a pen 3 Ensure that there are no kinks in the pipeworkdown the side of the ma...

Page 21: ...chnical support Before calling Please ensureyouhavetheserial number and model of equipment Please record details of the results of the recommended tests Notes pages are provided at the back of this us...

Page 22: ...PARTS LIST...

Page 23: ...e user damage D That PremiumLift s cleaning disinfecting guidelines have beenfollowed E PremiumLift smaintenance guidelines havebeenfollowed please refertothe product manual F ThatALL maintenance hasb...

Page 24: ...24 www premiumlift com au NOTES...

Page 25: ...25 NOTES...

Page 26: ...26 www premiumlift com au NOTES...

Page 27: ...au All parts and data continually undergo further development and may therefore deviate from the details given Copyright PremiumLift Our Full Terms Conditions are available by request or can be found...

Reviews: