Cycle/Alternating Time

 

The cycle time can be selected by pressing the button on the control panel. The 
cycle time options are 6, 9, 12 or 25 minutes. Consult a medical professional for 
cycle times above ten minutes. The pump always start with 6min cycle.

 

Pressure Range

 

Pressure range digital display 30-80 mmHg 

 

Alarm Mute

 

The pump is fitted with an audible and visual alarm. Press the 

‘Mute’

 button to 

reset the alarm. Initially the alarm will sound after 40 mins if the mattress has not 

fully inflated while in the start-up mode. Once safely inflated the alarm will sound if 

it detects a cycle fault or low pressure fault on the system.

 

Alternating Mode

 

When pump is switched on for the first time it will inflate all the cells.

 

The alternating mode LED will illuminate to indicate that the pump is in low-pressure 
mode. 

Once all the cells have been inflated the alternating mode will activate (normal 

pressure  mode). The cells in the  mattress will then inflate and deflated to prove 

pressure relief for the service user.

 

Static Mode

 

The 

‘Static

 

Mode’

 can be operated by the user to inflate all the cells for patient care 

and transfers. Press the 

‘Cycle Time” button until all 4 LED will illuminate green for 

“Static Mode”

 

Maxi-Inflate Mode

 

Press the 

‘Maxi

-Inflate

 button to automatically inflate the mattress to its maximum 

pressure. This will take approximately 20 mins. During normal pressure operations the 
pump will monitor the pressures. 

 

Comfort Level (kg)

 

The pump has a scale printed on the front panel in kilograms. Press the comfort 

control button (-) or (+) until the nearest desired weight setting is illuminated on 
the display panel. When the mattress is in use the user can adjust the comfort 

setting as required for the service user’s needs. A slight increase of the comfort 

setting may be required if the service user needs more support when the bed is 
profiled. When the bed platform is laid flat readjust the setting as required for the 
service user.

 

Lock Button

 

Press “Lock” to lock pump panel.

 

Press “

Unl

ock” to 

unlock pump panel.

 

Bed Hooks (Back of Pump)

 

Attach the pump to the foot end of the bed. Adaptor brackets are available to fit 

different sizes of beds. If the pump cannot be attached to the foot end of the bed it is 
the 

user’s responsibility to ensure that the pump is placed in such a position that it will 

not cause harm or injury to the service user. 

The pump unit must be placed so 

that it can be clearly seen and monitored at all times.

 

 

OVERVIEW

 

Summary of Contents for AirCare Max-Pro ACMaxPP

Page 1: ...AirCare Max Pro Digital Pump User Manual...

Page 2: ...2www premiumlift com au...

Page 3: ...AFETY INSTRUCTIONS 3 2 SAFETY INFORMATION FOR THE OPERATOR 3 3 SAFETY INFORMATION FOR THE USER 3 4 SYMBOLS USED 3 5 CLEANING DISINFECTION 3 6 SERVICING MAINTENANCE 3 7 SERVICE LIFE DISPOSAL TECHNICAL...

Page 4: ...the therapy may be reduced All hoses must be free of kinks twists and must be properly connected and positioned so as not to cause any obstruction Do not position the pump in a way that prevents acce...

Page 5: ...found on page 15 Suitableforcontinuous use Do not modify the mattress cushion or pump unit in any way Do not connect to any other medical device or equipment Not for use in an oxygen enriched environ...

Page 6: ...g pressure surface designed to reduce surface pressure on the skin and therefore reduce risk of the development of pressure ulcers in susceptable patients This pump has audible and visual alarms for l...

Page 7: ...ltoftheirvocational training or briefing are entitled to deliver assemble disassemble and transport the product 2 NON COMPLIANT USE All uses deviating from the intended purpose which may also be hazar...

Page 8: ...ct each user in the safe operation of this system before it is put into service for the first time Advise the user of any hazards which may occur if not handled correctly Only persons who have been pr...

Page 9: ...ccording to EN 60601 Handle with care This way up Keep dry Recycling symbol Refers to packaging that can be recycled cardboard Fragile handle with care This product must be disposed of in a designated...

Page 10: ...cts damageorsignsofwearmustberectifiedwithoutdelay Onlyoriginal spare parts from PremiumLift may be used otherwise all guarantees or warranties will be invalidated See pages 16 17 for Routine Maintena...

Page 11: ...6 9 12 25 minutes 345 x 115 x 185mm 3 5 kg 14 lpm Manual PUMP POWER REQUIREMENTS Power Rating Voltage Fuse MedicalClassification Safety Standards 13W AC 240V 50Hz T 1AH 250V Medical Device Class 1 EN...

Page 12: ...Pipes from the mattress connect here The pump must be able to be disconnected in an emergency for CPR to be administered B Power Switch The power switch is on the left hand side of the pump When press...

Page 13: ...ansfers Press the Cycle Time button until all 4 LED will illuminate green for Static Mode H Maxi Inflate Mode Press the Maxi Inflate button to automatically inflate the mattress to its maximum pressur...

Page 14: ...ug from the mains supply TRANSPORT MODE If the pump needs to be disconnected for any reason attach the transport cap to the feed tubes connector This will prevent deflation retaining the remaining air...

Page 15: ...s mattressesshouldbethoroughlyrinsedwithcleanwatertoremoveany residues This will help prevent any long term compatibility issues associated with disinfectant residues Alternatively disinfectionmay bea...

Page 16: ...ed clean with a mild antiseptic solution All cells are replaceable and can besourced from PremiumLift POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance r...

Page 17: ...are Max Pro pump should be serviced every 12 months by PremiumLift approved personnel using genuine PremiumLift spare parts Failure to follow the service schedule may invalidate future warranty claims...

Page 18: ...e mains plug is not blown 4 The wall socket that the pump is connected to is working correctly 1 Connect the pump to the nearest working mainsoutlet 2 Replace the plug fuse with the correct fuse 3 Try...

Page 19: ...is not trapped in the bed mechanism Contact PremiumLift technical support Before calling Pleaseensureyouhavetheserial number and model of equipment 2 The CPR valve is not trapped and is in the closed...

Page 20: ...not appear to be alternating 1 Check that the static mode is not turned on 2 Carefully mark one ofthe inflated cells with a pen 3 Ensure that there are no kinks in the pipeworkdown the side of the ma...

Page 21: ...chnical support Before calling Please ensureyouhavetheserial number and model of equipment Please record details of the results of the recommended tests Notes pages are provided at the back of this us...

Page 22: ...PARTS LIST...

Page 23: ...e user damage D That PremiumLift s cleaning disinfecting guidelines have beenfollowed E PremiumLift smaintenance guidelines havebeenfollowed please refertothe product manual F ThatALL maintenance hasb...

Page 24: ...24 www premiumlift com au NOTES...

Page 25: ...25 NOTES...

Page 26: ...26 www premiumlift com au NOTES...

Page 27: ...au All parts and data continually undergo further development and may therefore deviate from the details given Copyright PremiumLift Our Full Terms Conditions are available by request or can be found...

Reviews: