13
www.premiumlift.com.au
C
Cycle/Alternating Time
The cycle time can be selected by pressing the button on the control panel. The
cycle time options are 6, 9, 12 or 25 minutes. Consult a medical professional for
cycle times above ten minutes. The pump always start with 6min cycle.
D
Pressure Range
Pressure range digital display 30-80 mmHg
E
Alarm Mute
The pump is fitted with an audible and visual alarm. Press the
‘Mute’
button to
reset the alarm. Initially the alarm will sound after 40 mins if the mattress has not
fully inflated while in the start-up mode. Once safely inflated the alarm will sound if
it detects a cycle fault or low pressure fault on the system.
F
Alternating Mode
When pump is switched on for the first time it will inflate all the cells.
The alternating mode LED will illuminate to indicate that the pump is in low-pressure
mode.
Once all the cells have been inflated the alternating mode will activate (normal
pressure mode). The cells in the mattress will then inflate and deflated to prove
pressure relief for the service user.
G
Static Mode
The
‘Static
Mode’
can be operated by the user to inflate all the cells for patient care
and transfers. Press the
‘Cycle Time” button until all 4 LED will illuminate green for
“Static Mode”
H
Maxi-Inflate Mode
Press the
‘Maxi
-Inflate
’
button to automatically inflate the mattress to its maximum
pressure. This will take approximately 20 mins. During normal pressure operations the
pump will monitor the pressures.
I
Comfort Level (kg)
The pump has a scale printed on the front panel in kilograms. Press the comfort
control button (-) or (+) until the nearest desired weight setting is illuminated on
the display panel. When the mattress is in use the user can adjust the comfort
setting as required for the service user’s needs. A slight increase of the comfort
setting may be required if the service user needs more support when the bed is
profiled. When the bed platform is laid flat readjust the setting as required for the
service user.
J
Lock Button
Press “Lock” to lock pump panel.
Press “
Unl
ock” to
unlock pump panel.
K
Bed Hooks (Back of Pump)
Attach the pump to the foot end of the bed. Adaptor brackets are available to fit
different sizes of beds. If the pump cannot be attached to the foot end of the bed it is
the
user’s responsibility to ensure that the pump is placed in such a position that it will
not cause harm or injury to the service user.
The pump unit must be placed so
that it can be clearly seen and monitored at all times.