FAULT

 

CHECK THAT

 

STAGE 2 CHECK

 

IF PROBLEM PERSISTS

 

The pump  appears 
to be running but  the

 

mattress is not inflating 

correctly and or  the 

low pressure light is 

illuminated.

 

PLEASE NOTE

 

Inflation can take up to 30

 

minutes.

 

1. The hoses are routed 

correctly (not kinked) and 
connected to the pump 
correctly.

 

1.  Disconnect and then re- 

connect the hoses to the outlet 
on the side of the pump.

 

 

2.  Open then reclose the CPR 

valve, make sure the valve 
is not trapped in the bed 
mechanism.

 

Contact PremiumLift technical 
support.

 

 

Before calling:

 

 

Please ensure you have the serial 
number and model of equipment.

 

 

2. The CPR valve is not trapped 

and is in the closed position.

 

 

Please record details of the results of 
the recommended tests. (Notes pages 
are provided at the back of this user 
manual).

 

1.  There are no leaks in the 

mattress.

 

 

2.  The tubes in the mattress are 

not disconnected or kinked.

 

 

 

1. Replace any damaged or 

leaking mattress parts with 
the correct genuine Harvest 
Healthcare spare parts.

 

 

 

2. Straighten out any kinked 

pipes and reconnect any 
disconnected joints.

 

 

 
 
 
 
 
 
 
 
 

 

19

 

w

w

w

.p

re

mi
u

ml
ift

.c

o

m.
a

u

 

 

19

 

19

20

 

Summary of Contents for AirCare Max-Pro ACMaxPP

Page 1: ...AirCare Max Pro Digital Pump User Manual...

Page 2: ...2www premiumlift com au...

Page 3: ...AFETY INSTRUCTIONS 3 2 SAFETY INFORMATION FOR THE OPERATOR 3 3 SAFETY INFORMATION FOR THE USER 3 4 SYMBOLS USED 3 5 CLEANING DISINFECTION 3 6 SERVICING MAINTENANCE 3 7 SERVICE LIFE DISPOSAL TECHNICAL...

Page 4: ...the therapy may be reduced All hoses must be free of kinks twists and must be properly connected and positioned so as not to cause any obstruction Do not position the pump in a way that prevents acce...

Page 5: ...found on page 15 Suitableforcontinuous use Do not modify the mattress cushion or pump unit in any way Do not connect to any other medical device or equipment Not for use in an oxygen enriched environ...

Page 6: ...g pressure surface designed to reduce surface pressure on the skin and therefore reduce risk of the development of pressure ulcers in susceptable patients This pump has audible and visual alarms for l...

Page 7: ...ltoftheirvocational training or briefing are entitled to deliver assemble disassemble and transport the product 2 NON COMPLIANT USE All uses deviating from the intended purpose which may also be hazar...

Page 8: ...ct each user in the safe operation of this system before it is put into service for the first time Advise the user of any hazards which may occur if not handled correctly Only persons who have been pr...

Page 9: ...ccording to EN 60601 Handle with care This way up Keep dry Recycling symbol Refers to packaging that can be recycled cardboard Fragile handle with care This product must be disposed of in a designated...

Page 10: ...cts damageorsignsofwearmustberectifiedwithoutdelay Onlyoriginal spare parts from PremiumLift may be used otherwise all guarantees or warranties will be invalidated See pages 16 17 for Routine Maintena...

Page 11: ...6 9 12 25 minutes 345 x 115 x 185mm 3 5 kg 14 lpm Manual PUMP POWER REQUIREMENTS Power Rating Voltage Fuse MedicalClassification Safety Standards 13W AC 240V 50Hz T 1AH 250V Medical Device Class 1 EN...

Page 12: ...Pipes from the mattress connect here The pump must be able to be disconnected in an emergency for CPR to be administered B Power Switch The power switch is on the left hand side of the pump When press...

Page 13: ...ansfers Press the Cycle Time button until all 4 LED will illuminate green for Static Mode H Maxi Inflate Mode Press the Maxi Inflate button to automatically inflate the mattress to its maximum pressur...

Page 14: ...ug from the mains supply TRANSPORT MODE If the pump needs to be disconnected for any reason attach the transport cap to the feed tubes connector This will prevent deflation retaining the remaining air...

Page 15: ...s mattressesshouldbethoroughlyrinsedwithcleanwatertoremoveany residues This will help prevent any long term compatibility issues associated with disinfectant residues Alternatively disinfectionmay bea...

Page 16: ...ed clean with a mild antiseptic solution All cells are replaceable and can besourced from PremiumLift POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance r...

Page 17: ...are Max Pro pump should be serviced every 12 months by PremiumLift approved personnel using genuine PremiumLift spare parts Failure to follow the service schedule may invalidate future warranty claims...

Page 18: ...e mains plug is not blown 4 The wall socket that the pump is connected to is working correctly 1 Connect the pump to the nearest working mainsoutlet 2 Replace the plug fuse with the correct fuse 3 Try...

Page 19: ...is not trapped in the bed mechanism Contact PremiumLift technical support Before calling Pleaseensureyouhavetheserial number and model of equipment 2 The CPR valve is not trapped and is in the closed...

Page 20: ...not appear to be alternating 1 Check that the static mode is not turned on 2 Carefully mark one ofthe inflated cells with a pen 3 Ensure that there are no kinks in the pipeworkdown the side of the ma...

Page 21: ...chnical support Before calling Please ensureyouhavetheserial number and model of equipment Please record details of the results of the recommended tests Notes pages are provided at the back of this us...

Page 22: ...PARTS LIST...

Page 23: ...e user damage D That PremiumLift s cleaning disinfecting guidelines have beenfollowed E PremiumLift smaintenance guidelines havebeenfollowed please refertothe product manual F ThatALL maintenance hasb...

Page 24: ...24 www premiumlift com au NOTES...

Page 25: ...25 NOTES...

Page 26: ...26 www premiumlift com au NOTES...

Page 27: ...au All parts and data continually undergo further development and may therefore deviate from the details given Copyright PremiumLift Our Full Terms Conditions are available by request or can be found...

Reviews: