FAULT
CHECK THAT
STAGE 2 CHECK
IF PROBLEM PERSISTS
The pump appears
to be running but the
mattress is not inflating
correctly and or the
low pressure light is
illuminated.
PLEASE NOTE
Inflation can take up to 30
minutes.
1. The hoses are routed
correctly (not kinked) and
connected to the pump
correctly.
1. Disconnect and then re-
connect the hoses to the outlet
on the side of the pump.
2. Open then reclose the CPR
valve, make sure the valve
is not trapped in the bed
mechanism.
Contact PremiumLift technical
support.
Before calling:
Please ensure you have the serial
number and model of equipment.
2. The CPR valve is not trapped
and is in the closed position.
Please record details of the results of
the recommended tests. (Notes pages
are provided at the back of this user
manual).
1. There are no leaks in the
mattress.
2. The tubes in the mattress are
not disconnected or kinked.
1. Replace any damaged or
leaking mattress parts with
the correct genuine Harvest
Healthcare spare parts.
2. Straighten out any kinked
pipes and reconnect any
disconnected joints.