ProtEX-MAX PD8-7000 Explosion-Proof Temperature Meter Instruction Manual
41
Display
Parameter
Action/Setting
u-rang
Underrange
Program mA output
for display
underrange
Break
Loop Break
Set relay condition if
loop break detected
mAx
Maximum
Program maximum
mA output allowed
min
Minimum
Program minimum
mA output allowed
Calib
Calibrate
Calibrate 4-20 mA
output (internal
reference source
used for scaling the
output)
4 mA
4 mA output
Enter mA output
value read by
milliamp meter with at
least 0.001 mA
resolution
20 mA
20 mA
output
Enter mA output
value read by
milliamp meter with at
least 0.001 mA
resolution
user
User I/O
Assign function keys
and digital I/O
F1
F1 function
key
Assign F1 function
key
F2
F2 function
key
Assign F2 function
key
Display
Parameter
Action/Setting
F3
F3 function
key
Assign F3 function
key
F4
F4 function
Assign F4 function
(digital input)
dI I
Digital input
1
Assign digital input 1-
4
dO 1
Digital
output 1
Assign digital output
1- 4
ICal
Internal
temp
calibration
Enter internal
temperature
calibration (used for
factory calibration
only)
mref
Measured
temp
reference
Enter the measured
reference
temperature in
degrees Celsius (e.g.
25.0)
0nmv
0 mV
Apply 0.000 mV input
Internal Calibration
100nmv
100 mV
Apply 100.000 mV
input
Diag
Diagnostics
Display parameter
settings
LED t
LED test
Test all LEDs
Info
Information
Display software and
S/N information
Erase
Erase
Erase MeterView Pro
software stored in
meter’s memory
Offset Adjust (
Adjust
)
This parameter allows the user to select an offset adjustment to the temperature being displayed. Offset
adjustment values can be either positive or negative and can be any number within
50.0
F (+27.8
C). The
offset adjustment value is programmed through the
Adjust
menu.
The offset adjustment feature can be useful to compensate for errors due to thermocouple junctions or
excessive lead wire resistance in RTDs.
Note: The offset adjustment value is not reset to zero when the type of temperature sensor is changed
(i.e. TC type or RTD curve). Celsius/Fahrenheit conversion must be made manually by the user.