ProtEX-MAX PD8-7000 Explosion-Proof Temperature Meter Instruction Manual
40
Advanced Features Menu
To simplify the setup process, functions not needed for most applications are located in the
Advanced
Features
menu.
Press and hold the Menu button for three seconds to access the advanced features of the meter.
Advanced Features Menu & Display Messages
The following table shows the functions and messages of the
Advanced Features
menu in the order they
appear in the menu.
Display
Parameter
Action/Setting
Adjust
Adjust
Set adjust value to
offset temperature
reading
T cal
T Cal
Enter temperature
calibration menu
T lo
T Low
Apply the low
temperature input
(e.g. 32
°
F)
T Hi
T High
Apply the high
temperature input
(e.g. 1000
°
F)
rtdtot
RTD Total
Set the number of
RTDs connected in
parallel to obtain
average temperature
filtEr
Filter
Set noise filter value
bypASs
Bypass
Set filter bypass value
Round
Round
Set the rounding
value for display
variables
seriAl
Serial
Set serial
communication
parameters
Display
Parameter
Action/Setting
SlaveId
Slave ID
Set slave ID or meter
address
baud
Baud rate
Select baud rate
Tr dLY
Transmit
delay
Set transmit delay for
serial communication
Parity
Parity
Select parity
Even, Odd, or None
with 1 or 2 stop bits
t-byt
Time byte
Set byte-to-byte
timeout
SelEct
Select
Enter the Select
menu
AoutPr
Analog
output
programmin
g
Program analog
output parameters
Source
Source
Select source for the
4-20 mA output
O-rang
Overrange
Program mA output
for display overrange