background image

PD6720 Field-Mount Loop-Powered Rate/Totalizer 

Instruction Manual 

 

15

 

Calibrating the Meter (

Cal

To scale the meter without a signal source  
refer to 

Scaling the Meter

 (

SCale

). 

The meter can be calibrated to display the process in 
engineering units by applying the appropriate input 
signal and following the calibration procedure. 
The use of a calibrated signal source is strongly  
recommended. 

 

1.  Press the 

Up 

arrow button to scroll to the 

Calibration

 menu (

CAL

) and press 

Enter

2.  The meter displays 

Inpt1

. Apply a known 

signal and press 

Enter

.  

The display will flash while accepting the 
signal. 

3.  After the signal is accepted, the meter displays 

dspl1

 Press 

Enter

.  

Enter a corresponding display value for the 
signal input, and press 

Enter

 to accept. 

4.  The meter displays 

Inpt2

. Apply a known 

signal and press 

Enter

.  

The display will flash while accepting the 
signal. 

5.  After the signal is accepted, the meter displays 

dspl2

. Press 

Enter

.  

Enter a corresponding display value for the 
signal input and press 

Enter

 to accept. 

After completing calibration the 

save?

 display will 

need to be acknowledged using the Enter key before 
calibration will take effect. 

 

Minimum Input Span 

The minimum input span is the minimum difference 
between input 1 and input 2 signals required to 
complete the calibration or scaling of the meter.  
The minimum span is 0.10 mA. 

If the minimum span is not maintained, the meter 
reverts to input 2, allowing the appropriate input 
signals to be applied. 
 

Re-Calibrating the Internal  
Calibration Reference (

ICaL

The 

Internal Calibration

 (

ICAL

) menu, located in the 

Advanced

 features menu, is used to recalibrate the  

internal calibration reference. Recalibration is  
recommended at least every twelve months. Refer to  

 

Input 

Signal 

Filter

 

(

FILTER

 

The noise filter is available for unusually noisy signals 
that cause an unstable process variable display. The 
noise filter averages the input signal over a certain 
period. The filter level can be set to low (

LO

), medium 

(

med

), high (

HI

), or off (

OFF

). The higher the filter 

setting, the longer the averaging time and so the 
longer the display may take to find its final value.  
The filter contains a noise filter bypass feature so that 
while small variations in the signal will be filtered out, 
large, abrupt changes to the input signal are 
displayed immediately. 
Internal Calibration (

ICAL

) page 

21

 for instructions. 

 

Setting the Time Base (

t.base

The meter calculates total based on rate and a time 
base of units per second, minute, hour, or day. 

Press the 

Enter 

button, at any time, to accept a 

setting or 

Menu

 button to exit without saving changes. 

 

 

Setting the Total Conversion Factor 
(

totCF

Total Conversion Factor is used to convert to a 
different unit of total display. For example, to display 

Summary of Contents for PD6720 Series

Page 1: ...Display Rate in Units per Second Minute Hour or Day Display Open Channel Flow with Programmable Exponent Feature Open Collector Pulse or Alarm Output Display Mountable at 0 90 180 270 SafeTouchTM Thro...

Page 2: ...Terms Conditions on www predig com for complete details 2020 Precision Digital Corporation All rights reserved Introduction The PD6720 is a rugged field mount loop powered rate totalizer fully featur...

Page 3: ...ime Base t base 15 Setting the Total Conversion Factor totCF 15 Manual or Automatic Total Reset Function t rST 16 Setting the Tag Display tag 16 Setting Up the Password PASSWRD 17 Locking the Meter 17...

Page 4: ...Enclosure Dimensions Side Cross Section View 7 Figure 3 Connector Board 8 Figure 4 Connections without Backlight 9 Figure 5 Connections with Loop Powered Backlight 9 Figure 6 Connections with Externa...

Page 5: ...Environmental Operating temperature range 40 to 75 C Storage temperature range 40 to 75 C Relative humidity 0 to 90 non condensing Connections Screw terminals accept 12 to 22 AWG wire Enclosure NEMA...

Page 6: ...able from 0 to 99 999 seconds Open Collector Output Rating Isolated open collector sinking NPN 30 VDC 150 mA max Alarm Output Assign to rate for high or low alarm trip point Assign to total for total...

Page 7: ...and set the display module aside Unpacking Remove the meter from box Inspect the packaging and contents for damage Report damages if any to the carrier If any part is missing or the meter malfunction...

Page 8: ...t from external supply OUTPUT NPN open collector output positive OUTPUT NPN open collector output negative RESET Contact closure alarm acknowledge pull up to 3 VDC RESET Contact closure alarm acknowle...

Page 9: ...ions are made to two terminals labeled RESET Connect to a contact closure source such as a relay or a pushbutton as shown in Figure 7 Figure 7 Reset Connections Open Collector Output Connections Outpu...

Page 10: ...sors are disabled when a mechanical button is pressed and will automatically be re enabled after 60 seconds of inactivity The SafeTouch buttons are designed to filter normal levels of ambient interfer...

Page 11: ...nds to access the Advanced features of the meter Right Reset Button Press the Right arrow button to move to the next digit or decimal position during programming Up Display Button Press the Up arrow b...

Page 12: ...ogram menu sCalE Scale Enter the Scale menu Cal Calibrate Enter the Calibrate menu Inpt1 Input 1 Calibrate input 1 signal or program input 1 value DspL1 Display 1 Program display 1 value Inpt2 Input 2...

Page 13: ...w to select next digit and Up arrow to increment digit The digit being changed blinks Press the Enter button at any time to accept a setting or MENU button to exit without saving changes The decimal p...

Page 14: ...functions are exclusive of each other The meter uses the last function programmed Only one of these methods can be employed at a time The Scale and Calibrate functions can use up to 32 points default...

Page 15: ...complete the calibration or scaling of the meter The minimum span is 0 10 mA If the minimum span is not maintained the meter reverts to input 2 allowing the appropriate input signals to be applied Re...

Page 16: ...on choose man dsabl To reset total upon total alarm set point choose auto enter a time delay t dly and proceed to programming the set point see Alarm Output alrm on page 19 Press the Enter button at a...

Page 17: ...returns to its password protected condition Disabling Password Protection To disable the password protection access the Password menu and enter the correct password twice as shown below The meter is n...

Page 18: ...total Set Set Point Program set point Reset Reset Point Program reset point Pulse Pulse Output Program pulse output Scaler Count MAX HZ Frequency Program pulse output maximum frequency Funct Function...

Page 19: ...deadband is one display count If set and reset points are programmed the same output will reset one count below set point Total alarm trip point program total set point Alarm reset is triggered by tot...

Page 20: ...rent source instead of using input scaling This process will override previously programmed scaling points Apply a live signal using a known accurate signal source InP 1 32 and then enter the correspo...

Page 21: ...ging time and so the longer the display may take to find its final value The filter contains a noise filter bypass feature so that while small variations in the signal will be filtered out large abrup...

Page 22: ...he factory defaults Instructions to load factory defaults Enter the Advanced features menu Press and hold Reset button when INFO is shown Press Enter when RESET DFALTS prompt is shown Note If Enter is...

Page 23: ...exit Max Min display readings Display response is too slow Check filter setting to see if it can be lowered to LO or OFF If the display locks up or the meter does not respond at all Perform hard rese...

Page 24: ...on or increment digit Cycle through maximum minimum and mA display mode ENTER Accept selection value and move to next selection Acknowledge alarms Max Min Mode While in Run Mode pressing Up Arrow will...

Page 25: ...990 www predig com LIM6720_C SFT049 Ver 1 200 up 07 20 Contact Precision Digital Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com Sales Support Call 800 34...

Reviews: