Loop Leader PD6622/4 Loop-Powered Rate/Totalizer Instruction Manual
12
Setup and Programming
The meter is factory calibrated prior to shipment to read in
milliamps. The calibration equipment is certified to NIST standards.
Overview
There are no jumpers to set; setup and programming is done through the front panel buttons.
The meter may be powered via the micro-USB connection located on the right side of the meter for the
purposes of programming only. The backlight will not work while the meter is powered via the USB con-
nection.
Front Panel Buttons and Status LED Indicators
PD6622
PD6624
Button
Symbol
Description
LCD
Status
Menu
!
Alarm Indicator
(PD6622 Only)
Right-Arrow/F1
Password Protection
Indicator
(PD6622 Only)
Up-Arrow/F2
G
Grand Total Indicator
(PD6622 Only)
Enter/F3
T
Total Indicator
(PD6622 Only)
Press the
Menu
button to enter or exit the
Programming Mode
at any time.
Press or hold the
Right-Arrow
button to scroll forward through the menus, select digits during nu-
meric programing, select characters during text programming, or decrement the value of a digit or
character selected with the
Up-Arrow
button.
Press and hold the
Right-Arrow
button to zero or clear digits/characters while in data-entry mode.
Press or hold the
Up-Arrow
button to scroll backwards through the menus or to increment the value
of a digit or character.
Press the
Enter