14.1 Notes on the Weighing mode
14.1.1 Set Weighing mode: FLOATINGDISPLAY
The value set for FLOATINGDISPLAY defines the period, after which
each new measurement is displayed.
For the definition of this period, the quality of the balance-location is
crucial. The stability control must also be suitably chosen.
Recommended values:
14.1.2 Set Weighing mode: Stability control
The value set for the stability control depends on the quality of the
balance location and must be correctly chosen in order to obtain
optimal, reproducible results. Choose:
• “HIGH STABILITY” at an
optimal
balance location,
• “MEDIUM STABILITY” at a
good
balance location or
• “LOW STABILITY” at a
critical
balance location
14.1.3 Set Weighing mode: Auto-Standby
The Auto-Standby mode turns off the balance automatically, if:
• the balance is tared and has shown “Zero” for at least 5 minutes
• the balance has received no remote control command via the inter-
• Optimal balance location:
FLOATINGDISPLAY 0.04 or
FLOATINGDISPLAY 0.08
• Good balance location:
FLOATINGDISPLAY 0.16
• Critical balance location:
FLOATINGDISPLAY 0.32