3 Starting up
22
3.9
Standardized balance
The standardized balances of the XB-Series are provided with the EU/
OIML licence or meet the local standardisation regulations.
The balance program and certain functions of the weighing output
differ from the standard program in the case of the standardized XB
balances – in accordance with the EU/OIML provisions.
3.10 Under-floor weighing
Objects which, because of their size or shape, cannot be put on the
scale, can be weighed by means of under-floor weighing.
Proceed as follows:
• Turn off the balance.
• Remove the scale-pan and the scale-pan support and turn the balan-
ce around.
• Push the metal-cover (1) on the floor of the balance to one side.
• Hang a small hook (available as an accessory, see Chapter 13 “Ac-
cessories”) into the aperture (3) of the now visible metal casting (2).
• Place the balance over an opening.
• Replace the scale-pan support and the scale-pan.
• Level the balance (see Chapter 3.6 “Levelling the balance”).
• Switch on the balance.
NOTE
If a circle appears in the main-display of a standardized balance, so
the indicated value is unstandardised.
In balances of class (1) the circle also stands for the warm-up
phase.
Your customer service will be happy to assist you at any time if you
have any questions on the standardization of the balance or on wor-
king with standardized balances of the XB-Series.