Appendix B - Measurement Specifications
85
External Current (Option E, current input)
DC
0.1% + 0.01A
0.02Hz to 20Hz range
0.1% + 0.002A + 0.1%/Hz
add 0.15% below 0.04Hz if AC coupled
add 0.01A for peak measurements
0.2Hz to 200Hz range
0.1% + 0.002A + 0.01%/Hz
add 0.05% below 0.4Hz if AC coupled
add 0.01A for peak measurements
2Hz to 2KHz range
0.1% + 0.002A + 0.001%/Hz
add 0.05% below 4Hz if AC coupled
add 0.01A for peak measurements
20Hz to 5KHz range
0.1% + 0.003A + 0.4%/KHz
add 0.05% below 40Hz if AC coupled
add 0.03A for peak measurements
20Hz to 75KHz range
0.1% + 0.005A + 0.02%/KHz
add 0.1% below 40Hz if AC coupled
add 0.05A for peak measurements
Phase (to Voltage)
0.1° + 0.05°/KHz
Phase (to other phase)
0.1° + 0.05°/KHz
Current to Current Crosstalk
>100dB decreasing linearly to >70dB at 75KHz
VA
(Applied Voltage) x Current Ac (Applied Current) x Voltage Accuracy
Power Factor
Combination of phase and crosstalk specifications
At 115 or 230Vrms (current >10% range for option fitted) -
PF =
1.0000
0.9000
0.7000
0.5000
0.3000
0.1000
0.0000
@ 50/60Hz
0.0001
0.0008
0.0013
0.0016
0.0017
0.0018
0.0018
@ 400Hz
0.0001
0.0009
0.0015
0.0018
0.0020
0.0021
0.0021
@ 30KHz
0.0005
0.0218
0.0318
0.0341
0.0329
0.0295
0.0273
@ 75KHz
0.0056
0.0725
0.1014
0.1068
0.1015
0.0897
0.0822
Summary of Contents for PA2553
Page 1: ...Powertek USER S GUIDE PA2553 PA2553E Three Phase Power Analyser ...
Page 2: ......
Page 34: ...34 PA2553 User s Guide ...
Page 50: ...50 PA2553 User s Guide ...
Page 64: ...64 PA2553 User s Guide ...
Page 78: ...78 PA2553 User s Guide ...