Calibration
73
4.
Wait one minute for the readings Aa to settle and press ACCEPT.
5.
Wait one minute for the readings to settle to 0
±
0.1 and press ACCEPT.
6.
Press NEXT when the readings are within specification. The following screen will
display.
Figure 104. Load "B" Current Cal Point display
7.
Move the positive lead of the calibrator to SOURCE B binding post.
8.
Move the calibrator negative lead to the LOAD B binding post.
9.
Repeat steps 3 through 6—apply proper current; wait for readings Ab to settle to 0
±
0.1 and press ACCEPT; wait another minute for readings to settle and press
ACCEPT. Press NEXT and the following screen will display.
Figure 105. Load "C" Current Cal Point display
10.
Move the calibrator positive lead to the SOURCE C binding post.
11.
Move the calibrator negative lead to the LOAD C binding post.
12.
Repeat steps 3 through 5—apply proper current; wait for readings Ac to settle to 0
±
0.1 and press ACCEPT; wait another minute for readings to settle and press
ACCEPT.