Current Calibration
The current cal point display will show one of the following prompts, depending on
option fitted—
INPUT = 2ADC SOURCE L TO LOAD L (8Apk Option)
INPUT = 10ADC SOURCE L TO LOAD L (40Apk Option)
Note:
The current level shown on the display depends upon which current input option
is fitted.
Figure 102. Internal Current Cal Point display
WARNING:
SHOCK HAZARD. LETHAL VOLTAGES OR CURRENT MAY BE PRESENT.
ENSURE NO VOLTAGE OR CURRENT EXISTS ON THESE CONNECTIONS PRIOR TO
ATTEMPTING TO CONNECT TO THESE INPUT TERMINALS.
♦
To perform the internal current cal point—
1.
Connect the positive lead of the calibrator to SOURCE A binding post.
2.
Connect the negative lead of the calibrator to the LOAD A binding post.
3.
Apply the proper current input as shown on the display.
Figure 103. Load "A" Current Cal Point display
!