Accuracy
Specified accuracy is valid after 10 minutes warm up in temperature range 23 ± 3
!
C and for power
load up to 1W. Uncertainties include long-term stability, temperature coefficient, linearity, load and line
regulation and traceability of factory to National calibration standards. Accuracies assigned in % are related
to the set value. Specified accuracy is one-year accuracy.
MC642 Resistance accuracy
Range / Resolution
Accuracy
100.000 m
- 200.000 m
0.05 % + 15 m
200.01 m
- 2.00000
0.05 % + 15 m
2.0001
- 20.0000
0.05 % + 15 m
20.001
- 200.000
0.05 % + 15 m
200.01
- 2000.00
0.02 %
2.0001 k
- 20.0000 k
0.02 %
20.001 k
- 200.000 k
0.02 %
0.20001 M
- 2.00000 M
0.02 %
2.0001 M
- 20.0000 M
0.05 %
Table 7 MC642 Resistance accuracy
MC642 Pt simulation accuracy
Temperature range
Pt10-Pt99
Pt100-Pt20000
-200.000…0.000
!
C
0.5
!
C
0.15
!
C
0.001…850.000
!
C
1.0
!
C
0.2
!
C
Table 8 MC642 Pt simulation accuracy
MC642 Ni simulation accuracy
Temperature range
Ni10-Ni99
Ni100-Ni20000
-60.000…300.00
!
C
0.4
!
C
0.1
!
C
Table 9 MC642 Ni simulation accuracy
MC642 Frequency response
R
Max. AC/DC difference
100 Hz
1 kHz
10 kHz
100 m
0.05 %
0.20 %
5.00 %
1
0.02 %
0.10 %
0.50 %
10
0.01 %
0.02 %
0.10 %
100
0.01 %
0.10 %
0.60 %
1 k
0.06 %
0.60 %
6.00 %
10 k
0.60 %
6.00 %
100 k
6.00 %
Table 10 MC642 Frequency response
Temperature coefficient
Temperature coefficient outside of the reference temperature range is 10 % of the stated specification per °C
(for example 2x specification for 36°C).
Power coefficient
Power coefficient above 1W is 50 % of the stated specification per W (for example 3x specification for 5W
load).
Short and Open simulation
„SHORT“ resistance is lower than 40 m
. Maximal allowed current is 500 mA.
„OPEN“ resistance is higher than 10 G
. Maximal allowed voltage is 200 Vpk.
Note: Resistance values in range 100 m
- 20 M
are calibrated absolutely. Resistance value is not defined
against SHORT position. Short and Open positions are intended for functional checking of tested instrument
only.
71
Operation manual
Summary of Contents for MC642
Page 1: ...MC642 Programmable Resistance Decade Operation manual ...
Page 2: ......
Page 4: ......