:CALibration:SECure:PASSword
Syntax:
:CALibration:SECure:PASSword <DNPD>
Description:
This command validates entered password and enables calibration access if verification is
successful. Access is invalidated after reset or if CAL:SEC:EXIT command is issued.
Calibration password can be changed from decade system menu
SETUP->Calibration-
>Change password.
Parameters:
<DNPD>
Range 0 ... 4294967295 (default 0)
Remarks:
Overlapped command
Example:
CAL:SEC:PASS 0
:CALibration:SECure:EXIT
Syntax:
:CALibration:SECure:EXIT
Description:
This command exits calibration mode and access.
Parameters:
None
Remarks:
Overlapped command
Example:
CAL:SEC:EXIT
:DISPlay:ANNotation:CLOCk:DATE:FORMat
Syntax:
:DISPlay:ANNotation:CLOCk:DATE:FORMat <CPD>
:DISPlay:ANNotation:CLOCk:DATE:FORMat?
Description:
This command sets format of date displayed on device screen.
Parameters:
<CPD>
{MDYS|MDYA|DMYS|DMYO|DMYA|YMDS|YMDO} (default MDYS)
·
MDYS
M/D/Y format
(M-month, D-day, Y-year)
·
MDYA
M-D-Y format
·
DMYS
D/M/Y format
·
DMYO
D.M.Y format
·
DMYA
D-M-Y format
·
YMDS
Y/M/D format
·
YMDO
Y.M.D format
Remarks:
Overlapped command
Value is not affected by reset or by *RST command
Example:
DISP:ANN:CLOC:DATE:FORM MDYS
DISP:ANN:CLOC:DATE:FORM? Response: MDYS