FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599102AV • 15 September 2020
Page 33 of 36
BACnet
Object Name
Object Name Description
Gas1 Disp
The displayed gas pressure for a gas
badge.
Gas2 Disp
Gas3 Disp
Gas4 Disp
Gas5 Disp
Gas6 Disp
Gas7 Disp
Gas8 Disp
G1HiSpSet
The gas high pressure alarm setpoint for a
gas badge.
G2HiSpSet
G3HiSpSet
G4HiSpSet
G5HiSpSet
G6HiSpSet
G7HiSpSet
G8HiSpSet
G1LoSpSet
The gas low pressure alarm setpoint for a
gas badge.
G2LoSpSet
G3LoSpSet
G4LoSpSet
G5LoSpSet
G6LoSpSet
G7LoSpSet
G8LoSpSet
G1Mismatc
Gas Mismatch Alarm Status for a setup gas.
A displayed value of 0= Normal, 1=Alarm.
G2Mismatc
G3Mismatc
G4Mismatc
G5Mismatc
G6Mismatc
G7Mismatc
G8Mismatc
G1HiAlarm
Gas Badge High Alarm Status. A displayed
value of 0=Normal, 1=Alarm.
G2HiAlarm
G3HiAlarm
G4HiAlarm
G5HiAlarm
G6HiAlarm
G7HiAlarm
G8HiAlarm
Appendix A. BACnet Object Names
BACnet
Object Name
Object Name Description
G1HiAlarm
Gas Badge High Alarm Status. A displayed
value of 0=Normal, 1=Alarm.
G2HiAlarm
G3HiAlarm
G4HiAlarm
G5HiAlarm
G6HiAlarm
G7HiAlarm
G8HiAlarm
G1LoAlarm
Gas Badge Low Alarm Status. A displayed
value of 0=Normal, 1=Alarm.
G2LoAlarm
G3LoAlarm
G4LoAlarm
G5LoAlarm
G6LoAlarm
G7LoAlarm
G8LoAlarm
G1Missing
Gas Transducer Missing Alarm Status. A
displayed value of 0=Normal, 1=Alarm.
G2Missing
G3Missing
G4Missing
G5Missing
G6Missing
G7Missing
G8Missing
B#R#Alarm
“Master Alarm input alarm status. A
displayed value of 0=Normal, 1=Alarm.
B# = Master Badge #
R# = Row # within a Master Alarm Badge
Ex: B1R2 = The 1st Master Alarm badge has
an alarm for input that was setup for the
2nd row.
(Note: For badges 10, 11, and 12 ‘Alarm’ will
show up as ‘Alar’ in the BACnet object name
due to spacing limits.)”
BattAlarm
“PLC/HMI battery alarm status. A displayed
value of 0=Normal, 1=Alarm.
(Note: To maintained saved settings,
change the battery while the HMI/PLC is still
powered by 24VDC power.)”
Gen Fault
General Fault status. A displayed value of
0=Normal, 1=Alarm.
Summary of Contents for AAP10-OAV
Page 2: ......