Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599102AV • 15 September 2020
Page 32 of 36
FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Specifications
Operating temperature range: -4°F to 131°F
Storage temperature range: -22°F to 158°F
AC Input: Dry Contacts Input capable of 24 VDC 120 VAC
DC output: Dry Contacts capable of 30 VDC or 120 VAC max output
Input fuse: 2 Amp
Power consumption: 30 W Maximum
Pressure measurement accuracy: All positive pressure transducers: 0-200 psi ±0.5% at full scale
All vacuum transducers: 0-30 inHg ±0.5% at full scale
Dimensions (Width X Height X Depth)
Rough-in box – small: 12.4” X 10.4” X 4”
Rough-in box – large: 12.8” X 13.7” X 4”
Front panel – Small: 14” X 11.8”
Front Panel – Large: 14” X 15.4”
Transducers: 1.18” Diameter X 2.03” tall (Does not include DISS fitting)
Summary of Contents for AAP10-OAV
Page 2: ......