Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599102AV • 15 September 2020
Page 16 of 36
FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Combo Alarm Panel Wiring Schematic
23 Digital Inputs and 5-6 Gas Combo Alarm Panels – AP000C236AJ
TRANS.
1
RE
D
BL
AC
K
TRANS.
2
TRANS.
3
TRANS.
4
RE
D
BL
AC
K
RE
D
BL
AC
K
RE
D
BL
AC
K
RED (4X)
Note:Transducers and transducer wiring is field installed
BLACK (4X)
All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or
critical branch of the essential electrical system as required in NFPA 99
BLUE-WHITE (2X)
BLUE-WHITE
PLC/HMI
OUTPUT BAR
BROWN (4X)
BLUE (9X)
BLUE (14X)
BLUE-W (2X), BLUE (9X), BROWN (4X)
HORN
+
-
BLUE-W, BLUE (14X)
BL
UE
-W
HI
TE
COM
3
OUT
0
COM
4
OUT
6
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
A0 A1 A2 A3
BLUE
BLUE
YELLOW
I
0
I
1
I2
I3
I4
I5
I6
I7
I8
I9
I10
I11
I12
I13
I14
I15
I16
I17
I18
I19
I20
I21
I22
FOR BUILDING MAIN POWER TIE IN
GND LINE N
I1I
I0I
BLUE-W (2X)
RED
BLACK
POWER
SUPPLY
120V AC
24V DC
xx WATTS
+ -
24V
DC
L N
CB
WHITE
BLUE
BLUE
BLUE
BLUE
BLUE
BLUE-WHITE
BLUE-WHITE
PLC/HMI
INPUT BAR 1 (BLACK)
+V
0V
BLUE-WHITE
GREEN
BLUE
PLC/HMI
BLUE-W (2X)
BROWN (4X)
PLC/HMI
INPUT BAR 2 (GRAY)
BLUE-WHITE
BLUE-W
BLUE (9X)
BLUE (14X)
BLUE
GREEN
BLUE-WHITE
15
IN
AI3
14
IN
AI2
13
COM
1
12
AI1
11
AI0
10
IN
8
9
IN
7
8
IN
6
7
IN
5
6
IN
4
5
IN
3
4
IN
2
3
IN
1
2
IN
0
1
COM
0
COM2
16
I9
17
I10
18
I11
19
I12
20
I13
21
I14
22
I15
23
I16
24
I17
25
I18
26
I19
27
I20
28
I21
29
I22
30
Transd. 6
Transd. 5
14
13
12
I3I
11
CM
3
10
9
I2I
8
CM
2
7
6
5
I1I
4
CM
1
3
2
I0I
1
CM
0
Transd. 6
Transd. 5
14
13
12
I3I
11
CM
3
10
9
I2I
8
CM
2
7
6
5
I1I
4
CM
1
3
2
I0I
1
CM
0
O
+