FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599102AV • 15 September 2020
Page 23 of 36
Alarm System.
The Area alarms all begin with Gas #(1 thru 8) and the
Master alarms all begin with Badge #(1 thru 12).
The default Area alarms for each individual gas are as
follows:
• Sensor Mismatch Alarm
• High Setpoint Alarm
• Low Setpoint Alarm
• Sensor Missing
The Master alarm digital input signals are grouped
together under one gas badge and are indicated by
the source gas (these are edited in the “Master Alarm
Settings” screen).
Up to 7 digital input signals are available for each
badge, or piece of source equipment.
The Alarm Status screen has the same sorting ability
as the Alarm History screen – just press the sort button
in the upper left corner.
Alarm Summary
The Alarm Summary screen shows a list of all active
alarm conditions.
This screen is also accessible from the red bar with
the scrolling marquee at the bottom of the screen that
pops up during an alarm event. Press the “list” icon,
which is second from the left on the red bar.
Alarm conditions can be acknowledged and cleared
on this screen. Further details are given in the “Alarm
Event” section under “Operation.”
Gas Settings
This screen is only applicable to Area alarms and
Combo alarms.
In the gas settings page, you are able to change
the medical gas type, add/subtract a gas, designate
the location monitored for each gas, and manually
change the high/low alarm settings.
NOTE:
All Powerex
FirstCall™ Area Alarms are pre-programmed from the
factory with the correct medical gases and default high/
low alarm settings.
Location where each gas is being used can be entered
in this screen as well as via the home screen.
To edit/add/delete a gas badge, first press the edit
button next to the row number on the left. This makes
that row active in the editing row at the top highlighted
in yellow.
The gas is edited in the first column, and can be
selected from a preloaded list of NFPA 99 medical gases.
The correct NFPA 99 color combination is automatically
selected when a gas is selected.
NFPA 99 Gas Color Combinations:
Oxygen
Instrument Air
O2 / He
Medical Air
Helium
Lab Air
Medical Vacuum
Surgical Air
Lab Vacuum
Nitrous Oxide
Argon
Oxygen Hyp.
Nitrogen
CO2 / O2
Medical Air Hyp.
Carbon Dioxide
O2 / CO2
Carbon Dioxide Hyp.
WAGD
He / O2
AGSS
The Unit of Measure (UOM) for each gas can be
changed. For all NFPA 99 applications, PSIG should
be selected for all positive pressure gases and IN HG
should be selected for Medical Vacuum, WAGD, and
Lab Vacuum.
High/Low settings default settings for each gas are as
follows:
Medical Gas
Low Setpoint
(PSI)
High Setpoint
(PSI)
Oxygen
44
66
Medical Air
44
66
Medical Vacuum
12 inHg
N/A
Nitrous Oxide
40
60
Nitrogen
140
200
Carbon Dioxide
40
60
WAGD
12 inHg
N/A
Instrument Air
140
200
Summary of Contents for AAP10-OAV
Page 2: ......